Configure a custom test as follows:
1.
On the last page of the SETUP mode, select a custom test name; then press
E.
If you want to edit the name of the custom test, press E; then use the
2.
alphanumeric display to enter a new name.
3.
To select a test standard, which determines the default values for your
custom standard, select the current standard under USE DEFAULT
VALUES FROM on page 1 of the custom test displays.
To see and change other parameters, return to page 1 of the CONFIGURE
4.
CUSTOM TEST displays; then press $ Page Down.
Fiber Test Reports
Fiber test reports can be saved, viewed, printed, and uploaded to a PC in the
same way as other DSP-4000 test reports. See Chapter 5 in the DSP-4000
Users Manual for more information.
Maintenance
Clean the case with a soft cloth dampened with water or a mild detergent. Do
not use abrasives or solvents.
Clean the optical connectors as described under "Cleaning the Fiber
Connections". Always cover the connectors with a dust cap when not in use.
Do not open the DSP-FTA410 case. There are no user-serviceable parts inside.
Calibration
Have the DSP-FTA410 units calibrated at an authorized Fluke Service Center
every 12 months.
Fiber Test Reports
39