4. Insert the ATK2 sample
Carefully grip the upper part of the ATK2 using the furnace tongs
(Caution: fracture risk of the ceramic) and insert it into the holes
designated for this purpose until it snaps into place.
5. Press on sample
If necessary, use the furnace tongs to apply slight pressure to the
centre of the calibration base until the calibration sample clicks into
place. Observe the corresponding markings.
6. Start the calibration
Press the START key to start the calibration program. The progress of
the calibration program is shown in the display.
7. Complete the calibration
The result is displayed at the end of the calibration.
Temperature calibration successful
Temperature calibration failed
The calibration value is the difference between the measured current
temperature and the desired set temperature.
At the end of the program, open the furnace head and carefully remove the ATK2 using the furnace tongs and place it on
the cooling tray to allow it to cool. Replace the firing plate using the furnace tongs.
To return to the home screen, press either the touch button [Home] in the navigation bar or the HOME key on
the membrane-sealed keypad.