5. Operation and Configuration
5.1.7 Optical status and progress display (OSD)
The Optical Status Display (OSD) (12) shows the most important statuses of the furnace. The following activities are
indicated:
Colour
Yellow
Yellow (flashing)
Green
Orange
Red
Blue
Progress display:
During an active process, the progress of the program is also shown by the OSD. This is done by stage-to-stage lighting up
of the lateral OSD (12b).
5.1.8 User code
For safety reasons, a user code is required for certain settings. The user code ex factory is as follows:
The user code can be individually changed. See Chapter 5.4 Settings for details.

5.2 Firing programs and programming options

5.2.1 Program structure
The furnace offers several types of programs:
a. Programs for Ivoclar Vivadent materials
b. 500 free, individually adjustable programs
b. 500 free, individually adjustable programs on a USB stick
The program types are subdivided into program groups. Each group consists of 20 programs.
All programs are equivalent and therefore full-fledge programs. For each program, all the parameters can be adjusted.
a) Programs for Ivoclar Vivadent materials (see enclosed Program Table)
When the furnace is delivered ex factory, the Ivoclar Vivadent programs already contain the recommended material
para meter settings and are write-protected. Consequently, it is not possible to accidentally overwrite the programs.
If required, however, the parameters can be changed and overwritten at any time, if the programs are to be used for other
purposes. Therefore, these programs are also available as individual programs.
In case of software updates, individually changed parameters in Ivoclar Vivadent programs might be reset to the
factory settings or changed!
b) Free, individually adjustable programs
The programs are designed in such a way that they can be either used as conventional, one-stage programs,
or as two-stage programs, if required. The programs and program groups can be individually named.
In case of software updates, individually changed parameters in free programs are NOT reset to the factory
settings or changed!
26
Activity
The furnace is performing the self-test or is not ready for use, as the recommended
temperature range for a program start has not yet been reached.
Information, note or error message
The furnace is ready for use; the currently selected program can be started.
The program is closing the furnace head or is in predrying mode.
The program is heating up or is in holding time mode.
The program is in long-term cooling or is opening the furnace head.