Extech Instruments 793950343524 User Manual

Extech Instruments 793950343524 User Manual

Oyster meter ph, conductivity, tds, salinity, and orp (mv)

Advertisement

Quick Links

USER GUIDE 
 
 
Oyster Meter 
pH, Conductivity, TDS, Salinity, and ORP (mV)  
 
Model 341350A  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1.800.561.8187
information@itm.com
www.
.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 793950343524 and is the answer not in the manual?

Questions and answers

Summary of Contents for Extech Instruments 793950343524

  • Page 1 USER GUIDE      Oyster Meter  pH, Conductivity, TDS, Salinity, and ORP (mV)     Model 341350A                                                                      ...
  • Page 2: Meter Description

    Introduction  Congratulations on your purchase of the Extech Oyster Series meter. This device measures pH,  Conductivity, TDS, ORP and Salinity. These meters are intended for routine laboratory and field  testing. Accurate measurements are provided in a battery operated, portable meter with a hinged  display that can be adjusted to any viewing angle. This device is shipped fully tested and calibrated  and, with proper use, will provide years of reliable service.    Meter Description  Front and Side panels  LCD display  Battery compartment  Power button  Up/Down arrow buttons  MODE/CAL button  UNIT/SET button  STORE/RECALL button  Connector, Conductivity probe   Connector, pH probe, ORP probe  Connector, Temperature probe  Connector, AC adaptor      9 10 11   Display  Mode indicators  Primary display  Data Hold indicator  Probe Renew indicator  Auto Power Off indicator  Low battery indicator  Memory status indicators  Calibration status indicators  Temperature display  Unit indicators  341350A-en-US_V1.3 8/15 1.800.561.8187 information@itm.com www.
  • Page 3: Operation

    Operation  Battery Installation/Replacement  Open the battery cover by inserting a small coin into the latch slot and pressing downward. The  cover will release in the direction of the arrow. Install the new battery and replace the cover.          pH Measurements  Press the POWER button to turn the meter ON (the meter will cycle through a self‐check).  Press the MODE/CAL button until the PH icon switches on.  Press the UNIT/SET button to select ºC or ºF.  Connect the pH electrode to the side pH BNC connector.  Adjust the ▲▼buttons to match the temperature of the solution (or use the external  temperature probe to do so).  If required, calibrate the electrode (see pH Calibration).  Place the electrode in the sample solution and read the pH value on the display.  When all measurements have been taken, unplug the electrode and rinse in distilled water;  then shake and store it in the wetting cap with 4.0 M KCL or in a pH 4 buffer solution.  Conductivity/TDS/Salt Measurements   Press the POWER button to turn the meter ON.  Press the MODE/CAL button until the COND, SALT or TDS icon is displayed.  Connect the conductivity electrode to the side conductivity connector.  If required, calibrate the electrode (see Conductivity Calibration).  Place the electrode in the sample solution and read the value on the display.  When finished, unplug the electrode and rinse in distilled water.   ORP (mV) Measurements  Press the POWER button to turn the meter ON.  Press the MODE/CAL button until the ORP icon is displayed.  Connect the ORP electrode to the side ORP BNC connector.  Place the electrode in the sample solution and read the mV value on the display.  When all measurements have been taken, unplug the electrode and rinse in distilled water.  Store the probe tip in the wetting cap in a 4.0 M KCL or in a pH 4 buffer solution.  341350A-en-US_V1.3 8/15 1.800.561.8187 information@itm.com www. .com...
  • Page 4: Temperature Units

    Data Memory  25 readings can be stored to, and recalled from, the internal memory.  Storing readings  With the reading on the display, momentarily press the STORE/RECALL button. The STORE  and HOLD display icons will appear on the LCD and the reading will freeze.   Momentarily press the STORE/RECALL button again to confirm and to return to the normal  operation mode.   Up to 25 readings can be stored in this way.  Recalling Readings  Press and hold the STORE/RECALL button. The RECALL icon and the memory location  number will appear, followed by the data stored in the displayed memory location.   Use the ▼ ▲buttons to scroll through the memory locations and to view the stored data.   Momentarily press the STORE/RECALL button to exit the Recall Readings mode. End will  appear in the display; the meter will then return to the normal measurement mode.  Clearing the memory  With the meter on, press and hold the ▼ and ▲ buttons for 2 seconds. The CLr icon will appear  on the display, indicating that the memory has been erased.  Temperature Units  Press the UNIT button to select ºF or ºC.  Manual Temperature Compensation  In the pH mode, the solution temperature can be measured and set by using the optional  external temperature probe or, when the probe is not used, by adjusting the temperature  display. Press the ▲▼buttons to set the temperature when the MTC icon is displayed.     The Conductivity probe has a built‐in temperature sensor so manual temperature compensation  does not apply for Conductivity measurements.  Auto Power off (APO)  The meter is equipped with an Automatic Power Off feature. The meter will power down after a  10‐minute period of inactivity. This feature can be disabled by pressing and holding the POWER  button for 2 seconds (the APO display icon indicates that this feature is enabled). After disabling  APO, the next time the meter is powered up APO will be re‐enabled.  Reset to Default settings  The meter can be reset to its factory default configuration:  With the meter OFF, press and hold the POWER and STORE/RECALL buttons simultaneously  until dFLt  rSt appears in the display.  Release the buttons and the meter will continue with a normal startup.  Default values include: Calibration values, Conductivity mode, ºC, MTC, APO and RENEW off    341350A-en-US_V1.3 8/15 1.800.561.8187 information@itm.com www.
  • Page 5 pH Calibration (1, 2, or 3 points)  A two point calibration with a buffer of 7 plus 4 or 10 (whichever is nearest to the expected  sample value) is always recommended. A one point calibration (choose the value closest to the  expected sample value) or a three point calibration is also valid. For best accuracy, always  calibrate at the sample temperature. Frequency of calibration is dependent on how often the  meter is used, the care of the electrode, and strength of the samples tested. Typically, it is  recommended that calibration be performed once a day; or before each use if the meter is not  used on a daily basis.  Connect the pH probe and place it into a pH7 buffer.  Press the POWER button to turn the meter on and press the MODE/CAL button until the PH  icon appears. (Note: disable the Auto Power Off feature to avoid an automatic power off  during calibration).  If the MTC icon appears next to the temperature display, press the ▼ or ▲ button to set  the temperature of the pH buffer. If the optional temperature probe is used, insert the  probe into the buffer solution.   Press and HOLD the MODE/CAL button until the display begins to flash ‘7.00’. The meter  automatically recognizes the buffer and calibrates itself to that value. At the end of the  calibration cycle, SA and End will briefly appear on the display and the meter will return to  the normal operating mode.  When a calibration is performed, the calibration icons  (pH10)  (pH7)  (pH4) will be  cleared (calibration data is not erased) and will be replaced when a successful calibration is  performed for each buffer within one power on cycle. These icons indicate what calibration  levels were last performed. They do not indicate how recently the calibration was  performed or if the calibrations are still valid.  Remove the electrode from the pH7 buffer, rinse in a rinse solution and insert it into the  pH4 or pH10 buffer solution.  Repeat step 4 for the second calibration point and then the third point if desired.  Note: If the output of the electrode falls outside predetermined limits, the meter will cancel the  calibration, indicate End, and flash the RENEW icon. This typically happens when the electrode  has aged and needs replacement.  Note: To avoid cross contamination, always rinse the electrode in a rinse solution, like distilled  water, when changing from one buffer or sample to the next buffer or sample.    ORP Calibration  The ORP electrode does not require calibration.    However, you can test the ORP probe’s accuracy using an ORP standard.    341350A-en-US_V1.3 8/15 1.800.561.8187 information@itm.com...
  • Page 6 Conductivity (TDS & Salt) Calibration  Conductivity accuracy verification should be performed on a periodic basis; once per month is  the recommended cycle for normal use. If calibration is required, a conductivity standard  solution must be obtained. The meter can be calibrated in any or all of the three ranges.  Standardizing solutions of 84µS/cm, 1413µS/cm or 12.88mS/cm (12,880µS/cm) are used for the  automatic calibration recognition procedure. No other calibration values are permitted.  Calibration is always done in conductivity mode. Since salinity and TDS values are calculated  from conductivity values, this procedure also calibrates the Salinity and TDS ranges.  Fill a sample cup with the standardizing solution.  Turn the meter ON and insert the electrode into the solution. Tap or move the electrode in  the sample to dislodge any air bubbles.  Press and hold the MODE/CAL button (approximately 2 seconds) until the main display  starts flashing.  The meter will automatically recognize and calibrate to the standardizing solution. At the  end of the calibration, the display will briefly indicate “SA”, “End” and then return to the  measurement mode.  Note:  The “SA” will not appear if the calibration fails.  The “range calibrated” alert will appear in the display for each range that is calibrated  during that power on cycle.  Low range, 84µS/cm  Medium range, 1413µS/cm  High range, 12.88mS/cm (12,880µS/cm)  Note: The meter allows for a 1, 2 or 3 point calibration. If calibration is done for more than one  point, the lowest value standard should be done first to obtain the best accuracy.      Temperature (pH and Conductivity) Offset Calibration  This procedure allows for error correction of the external temperature probe (pH) or the  conductivity probe’s built‐in temperature sensor.  Switch to pH or Conductivity (Salt/TDS) mode.  Place the temperature probe or conductivity cell in the sample and allow the temperature  reading to stabilize.  Press and hold the UNIT/SET button until the ºC or ºF icon begins flashing.  Adjust the ▼ or ▲ buttons to set the display to indicate the known temperature of the  sample.  Momentarily press the UNIT/SET button to save the change and return to the measurement  mode.    341350A-en-US_V1.3 8/15 1.800.561.8187 information@itm.com www.
  • Page 7 pH Troubleshooting Chart  Symptom  Cause  Recommended Solution  Long response  Clogged Junction  Soak in 4.07 M KCL @ 60 C for 30  time or reading  minutes  drift  Oil, paint, dyes, suspended solids on  Rinse electrode alternately with  sensor  materials solvent then buffer 7.00  Dry Bulb  Long term storage without wetting   Soak electrode tip in wetting cap filled  with 1ml 7.00 buffer for 24 to 48 hours  Static Charge  Wiping electrodes  Rinse electrode in 7.0 buffer and blot.  Do no wipe electrode.  Same readings in  Cracked or broken bulb  Replace electrode. Use bulb guard.  different buffers  Avoid plunging electrode to bottom of  and samples  container and spinning bars. A wetting  cap will protect bulb between  measurements.  Erratic LCD  Samples have low ionic strength  For each 50 ml of sample add 1 drop  display  (lacks salt); e.g. distilled, de‐ionized,  (50uL) of SAT.KCL No alteration in pH  boiled, lake water (high pressure)  will occur by inert KCL.   ...
  • Page 8: Specifications

    SPECIFICATIONS    Ranges  Resolution  Accuracy  pH  0.00 to 14.00pH  0.01pH  ± 0.02pH  Conductivity  0.1µS  0.0 to 200.0µS  1µS  200 to 2000µS  ± 2% FS  2.00 to 20.00mS  0.01mS  TDS  0.0 to 134.0ppm  0.1ppm  134 to 1340ppm  1ppm  (calculated from Conductivity)  1.34 to 13.40ppt  0.01ppt  Salinity  0.0 to 100.0ppm  0.1ppm  (calculated from Conductivity)  100 to 1000ppm  1ppm  1.00 to 10.00ppt  0.01ppt  ORP  ‐1500 to 1500mV  1mV  ± 3mV  Temperature  32 to 194ºF  0.1º ≤ 99.9º  ± 2ºF/1ºC (meter + probe) ...

This manual is also suitable for:

341350a-p341350aOyster series

Table of Contents