Extech Instruments 793950343524 User Manual page 7

Oyster meter ph, conductivity, tds, salinity, and orp (mv)
Table of Contents

Advertisement

1.800.561.8187
pH Troubleshooting Chart 
Symptom 
Long response 
Clogged Junction 
time or reading 
drift 
Oil, paint, dyes, suspended solids on 
sensor 
Dry Bulb 
Long term storage without wetting   Soak electrode tip in wetting cap filled 
Static Charge 
Wiping electrodes 
Same readings in 
Cracked or broken bulb 
different buffers 
and samples 
Erratic LCD 
Samples have low ionic strength 
display 
(lacks salt); e.g. distilled, de‐ionized, 
boiled, lake water (high pressure) 
 
Notes on pH measurements and electrodes 
1.
The Electrode should be stored in its wetting cap until used. Use a pH 4 buffer solution or a 
4.0 M KCL solution. 
2.
If bubbles are seen in the bulb area, hold the electrode by its cap and shake downwards 
until bubbles are removed. 
3.
To improve speed of response, vigorously stir the electrode in the sample, buffer, or rinse 
solution. 
4.
After exposure to a sample, buffer, or rinse solution, shake the electrode with a snap 
motion to remove residual drops of solution. 
5.
When possible, use part of the next sample/buffer to be measured as a rinse solution,. 
6.
Keep buffers and samples at the same temperature to avoid temperature effects. 
7.
pH readings stabilize faster in some solutions than others; allow time to stabilize. 
8.
Electrodes deteriorate over time. If accuracy falls to 10% the electrode should be cleaned. If 
no improvement is observed, replace the electrode. 
Notes on the Conductivity/TDS/Salt Cell 
1.
Cell Storage: On sheathed cells, replace the sheath over the cell when storing. For non‐
sheathed versions, soak the cell tip in de‐ionized water for storage. 
2.
Cell Cleaning: After each use, the cell tip should be rinsed with de‐ionized water. If solids 
build up inside the cell carefully remove with a cotton swab soaked in solvent taking care 
not to touch the metal parts of the inner cell. 
www.
Cause 
Recommended Solution 
Soak in 4.07 M KCL @ 60
minutes 
Rinse electrode alternately with 
materials solvent then buffer 7.00 
with 1ml 7.00 buffer for 24 to 48 hours 
Rinse electrode in 7.0 buffer and blot. 
Do no wipe electrode. 
Replace electrode. Use bulb guard. 
Avoid plunging electrode to bottom of 
container and spinning bars. A wetting 
cap will protect bulb between 
measurements. 
For each 50 ml of sample add 1 drop 
(50uL) of SAT.KCL No alteration in pH 
will occur by inert KCL. 
7
.com
o
C for 30 
341350A-en-US_V1.3 8/15
information@itm.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 793950343524 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

341350a-p341350aOyster series

Table of Contents