Extech Instruments 341350A User Manual
Extech Instruments 341350A User Manual

Extech Instruments 341350A User Manual

Oyster meter ph, conductivity, tds, salinity, and orp (mv)
Hide thumbs Also See for 341350A:

Advertisement

Quick Links

 
 
Oyster Meter 
pH, Conductivity, TDS, Salinity, and ORP (mV)  
 
Model 341350A  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
GlobalScientificSupply
w.
GLOBAL SCIENTIFIC SUPPLY
.com
The Laboratory Supply Company!
USER GUIDE 
info@GlobalScientificSupply.co

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 341350A and is the answer not in the manual?

Questions and answers

Summary of Contents for Extech Instruments 341350A

  • Page 1 USER GUIDE      Oyster Meter  pH, Conductivity, TDS, Salinity, and ORP (mV)     Model 341350A                                                                    ...
  • Page 2: Meter Description

    LCD display  Battery compartment  Power button  Up/Down arrow buttons  MODE/CAL button  UNIT/SET button  STORE/RECALL button  Connector, Conductivity probe   Connector, pH probe, ORP probe  Connector, Temperature probe  Connector, AC adaptor      9 10 11   Display  Mode indicators  Primary display  Data Hold indicator  Probe Renew indicator  Auto Power Off indicator  Low battery indicator  Memory status indicators  Calibration status indicators  Temperature display  Unit indicators  341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 3: Operation

          pH Measurements  Press the POWER button to turn the meter ON (the meter will cycle through a self‐check).  Press the MODE/CAL button until the PH icon switches on.  Press the UNIT/SET button to select ºC or ºF.  Connect the pH electrode to the side pH BNC connector.  Adjust the ▲▼buttons to match the temperature of the solution (or use the external  temperature probe to do so).  If required, calibrate the electrode (see pH Calibration).  Place the electrode in the sample solution and read the pH value on the display.  When all measurements have been taken, unplug the electrode and rinse in distilled water;  then shake and store it in the wetting cap with 4.0 M KCL or in a pH 4 buffer solution.  Conductivity/TDS/Salt Measurements   Press the POWER button to turn the meter ON.  Press the MODE/CAL button until the COND, SALT or TDS icon is displayed.  Connect the conductivity electrode to the side conductivity connector.  If required, calibrate the electrode (see Conductivity Calibration).  Place the electrode in the sample solution and read the value on the display.  When finished, unplug the electrode and rinse in distilled water.   ORP (mV) Measurements  Press the POWER button to turn the meter ON.  Press the MODE/CAL button until the ORP icon is displayed.  Connect the ORP electrode to the side ORP BNC connector.  Place the electrode in the sample solution and read the mV value on the display.  When all measurements have been taken, unplug the electrode and rinse in distilled water.  Store the probe tip in the wetting cap in a 4.0 M KCL or in a pH 4 buffer solution.  341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 4: Manual Temperature Compensation

    Use the ▼ ▲buttons to scroll through the memory locations and to view the stored data.   Momentarily press the STORE/RECALL button to exit the Recall Readings mode. End will  appear in the display; the meter will then return to the normal measurement mode.  Clearing the memory  With the meter on, press and hold the ▼ and ▲ buttons for 2 seconds. The CLr icon will appear  on the display, indicating that the memory has been erased.  Temperature Units  Press the UNIT button to select ºF or ºC.  Manual Temperature Compensation  In the pH mode, the solution temperature can be measured and set by using the optional  external temperature probe or, when the probe is not used, by adjusting the temperature  display. Press the ▲▼buttons to set the temperature when the MTC icon is displayed.     The Conductivity probe has a built‐in temperature sensor so manual temperature compensation  does not apply for Conductivity measurements.  Auto Power off (APO)  The meter is equipped with an Automatic Power Off feature. The meter will power down after a  10‐minute period of inactivity. This feature can be disabled by pressing and holding the POWER  button for 2 seconds (the APO display icon indicates that this feature is enabled). After disabling  APO, the next time the meter is powered up APO will be re‐enabled.  Reset to Default settings  The meter can be reset to its factory default configuration:  With the meter OFF, press and hold the POWER and STORE/RECALL buttons simultaneously  until dFLt  rSt appears in the display.  Release the buttons and the meter will continue with a normal startup.  Default values include: Calibration values, Conductivity mode, ºC, MTC, APO and RENEW off    341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 5 If the MTC icon appears next to the temperature display, press the ▼ or ▲ button to set  the temperature of the pH buffer. If the optional temperature probe is used, insert the  probe into the buffer solution.   Press and HOLD the MODE/CAL button until the display begins to flash ‘7.00’. The meter  automatically recognizes the buffer and calibrates itself to that value. At the end of the  calibration cycle, SA and End will briefly appear on the display and the meter will return to  the normal operating mode.  When a calibration is performed, the calibration icons  (pH10)  (pH7)  (pH4) will be  cleared (calibration data is not erased) and will be replaced when a successful calibration is  performed for each buffer within one power on cycle. These icons indicate what calibration  levels were last performed. They do not indicate how recently the calibration was  performed or if the calibrations are still valid.  Remove the electrode from the pH7 buffer, rinse in a rinse solution and insert it into the  pH4 or pH10 buffer solution.  Repeat step 4 for the second calibration point and then the third point if desired.  Note: If the output of the electrode falls outside predetermined limits, the meter will cancel the  calibration, indicate End, and flash the RENEW icon. This typically happens when the electrode  has aged and needs replacement.  Note: To avoid cross contamination, always rinse the electrode in a rinse solution, like distilled  water, when changing from one buffer or sample to the next buffer or sample.    ORP Calibration  The ORP electrode does not require calibration.    However, you can test the ORP probe’s accuracy using an ORP standard.    341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 6 Press and hold the MODE/CAL button (approximately 2 seconds) until the main display  starts flashing.  The meter will automatically recognize and calibrate to the standardizing solution. At the  end of the calibration, the display will briefly indicate “SA”, “End” and then return to the  measurement mode.  Note:  The “SA” will not appear if the calibration fails.  The “range calibrated” alert will appear in the display for each range that is calibrated  during that power on cycle.  Low range, 84µS/cm  Medium range, 1413µS/cm  High range, 12.88mS/cm (12,880µS/cm)  Note: The meter allows for a 1, 2 or 3 point calibration. If calibration is done for more than one  point, the lowest value standard should be done first to obtain the best accuracy.      Temperature (pH and Conductivity) Offset Calibration  This procedure allows for error correction of the external temperature probe (pH) or the  conductivity probe’s built‐in temperature sensor.  Switch to pH or Conductivity (Salt/TDS) mode.  Place the temperature probe or conductivity cell in the sample and allow the temperature  reading to stabilize.  Press and hold the UNIT/SET button until the ºC or ºF icon begins flashing.  Adjust the ▼ or ▲ buttons to set the display to indicate the known temperature of the  sample.  Momentarily press the UNIT/SET button to save the change and return to the measurement  mode.    341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 7 Samples have low ionic strength  For each 50 ml of sample add 1 drop  display  (lacks salt); e.g. distilled, de‐ionized,  (50uL) of SAT.KCL No alteration in pH  boiled, lake water (high pressure)  will occur by inert KCL.    Notes on pH measurements and electrodes  The Electrode should be stored in its wetting cap until used. Use a pH 4 buffer solution or a  4.0 M KCL solution.  If bubbles are seen in the bulb area, hold the electrode by its cap and shake downwards  until bubbles are removed.  To improve speed of response, vigorously stir the electrode in the sample, buffer, or rinse  solution.  After exposure to a sample, buffer, or rinse solution, shake the electrode with a snap  motion to remove residual drops of solution.  When possible, use part of the next sample/buffer to be measured as a rinse solution,.  Keep buffers and samples at the same temperature to avoid temperature effects.  pH readings stabilize faster in some solutions than others; allow time to stabilize.  Electrodes deteriorate over time. If accuracy falls to 10% the electrode should be cleaned. If  no improvement is observed, replace the electrode.  Notes on the Conductivity/TDS/Salt Cell  Cell Storage: On sheathed cells, replace the sheath over the cell when storing. For non‐ sheathed versions, soak the cell tip in de‐ionized water for storage.  Cell Cleaning: After each use, the cell tip should be rinsed with de‐ionized water. If solids  build up inside the cell carefully remove with a cotton swab soaked in solvent taking care  not to touch the metal parts of the inner cell.  341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 8: Specifications

    TDS conversion ratio  0.67 (fixed)  Salinity conversion ratio  0.5 (fixed)  Auto Power OFF  After 10 minutes of inactivity (can be disabled)  Over range indication   “OL”  Operating Temperature  41ºF to 104ºF (5ºC to 40ºC)  Storage Temperature  ‐4ºF to 140ºF (‐20ºC to 60ºC)  Operating Humidity   Max 80% up to 87ºF (31ºC) decreasing linearly to 50% at  104ºF (40ºC)  Storage Humidity  <80%   Operating Altitude  7000’ (2000m) maximum  Power  9V alkaline battery or AC adapter  Dimensions  4.7 x 3.8 x 1.8” (118 x 96 x 45mm) with meter closed  Weight  12oz. (340g)        341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 9: Warranty

    Calibration, Repair, and Customer Care Services  FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we  sell. NIST certification for most products is also provided. Call the Customer Service Department for  information on calibration services available for this product. Annual calibrations should be  performed to verify meter performance and accuracy. Technical support and general customer  service is also provided, refer to the contact information provided below.  Support Lines: U.S. (877) 439‐8324; International: +1 (603) 324‐7800  Technical Support: Option 3; E‐mail: support@extech.com  Repair & Returns: Option 4; E‐mail: repair@extech.com  Product specifications are subject to change without notice  Please visit our website for the most up‐to‐date information  www.extech.com  FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA   ISO 9001 Certified    Copyright © 2013‐2015 FLIR Systems, Inc.  All rights reserved including the right of reproduction in whole or in part in any form  www.extech.com  341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 10: Garantie

    (AR) doit être délivré avant tout retour de produit. L’expéditeur prend à sa charge les frais d’expédition, le fret, l’assurance et  l’emballage correct de l’appareil afin de prévenir toute détérioration durant le transport. Cette garantie ne s’applique pas aux  dommages imputables à l’utilisateur, tels que l’usage impropre ou abusif, un mauvais câblage, une utilisation non conforme aux  spécifications, un entretien ou une réparation incorrecte, ou toute modification non autorisée. FLIR Systems, Inc. déclinera  spécifiquement toute garantie ou qualité marchande ou aptitude à l’emploi prévu, et ne sera en aucun cas tenu responsable  pour tout dommage conséquent, direct, indirect ou accidentel. La responsabilité totale de FLIR est limitée à la réparation ou au  remplacement du produit. La garantie définie ci‐dessus est inclusive et aucune autre garantie, écrite ou orale, n’est exprimée ou    implicite. Calibrage, réparation et services après‐vente  FLIR Systems, Inc. offre des services de calibrage et de réparation pour les produits Extech  Instruments que nous commercialisons. Nous fournissons également une certification NIST pour la  plupart des produits. Contactez notre service client pour toute information sur les services de  calibrage disponibles pour ce produit. Un calibrage doit être effectué chaque année pour vérifier  les performances et la précision du mètre. Nous offrons également une assistance technique et un  service à la clientèle. Veuillez vous reporter aux coordonnées fournies ci‐dessous.  Lignes d’assistance: États‐Unis (877) 439‐8324; international: +1 (603) 324‐7800  Service d’assistance technique : Option 3 ; E‐mail : support@extech.com  Réparations et retours : Option 4 ; E‐mail : repair@extech.com  Les spécifications produit sont sujettes à modifications sans préavis.  Pour les toutes dernières informations, veuillez visiter notre site Web.  www.extech.com  FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA  Certifié ISO 9001    Copyright © 2013‐2015 FLIR Systems, Inc.  Tous droits réservés, y compris la reproduction partielle ou totale sous quelque forme que ce soit.  www.extech.com  341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!
  • Page 11 Servicios de calibración, reparación y atención a clientes  FLIR Systems, Inc., ofrece servicios de reparación y calibración para los productos que vendemos  de Extech Instruments. Además ofrecemos certificación NIST para la mayoría de los productos.  Llame al Departamento de Servicio al Cliente para solicitar información de calibración para este  producto. Para verificar el funcionamiento y precisión se debe realizar la calibración anual. Además  se provee Soporte Técnico y servicios generales al cliente, consulte la información de contacto en  seguida.  Líneas de soporte: EE.UU. (877) 439‐8324; Internacional: +1 (603) 324‐7800  Soporte Técnico Opción 3; correo electrónico: support@extech.com  Reparación / Devoluciones: Opción 4; correo electrónico: repair@extech.com  Las especificaciones del producto están sujetas a cambios sin aviso  Por favor visite nuestra página en Internet para la información más actualizada  www.extech.com  FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA  Certificado ISO 9001  Copyright © 2013‐2015 FLIR Systems, Inc.    Reservados todos los derechos, incluyendo el derecho de reproducción total o parcial en cualquier medio www.extech.com    341350A-en-US_V1.3 8/15 GlobalScientificSupply GLOBAL SCIENTIFIC SUPPLY info@GlobalScientificSupply.co .com The Laboratory Supply Company!

Table of Contents