Model 34135-A measures pH, Conductivity and TDS. The ‘A’ indicates ‘meter-only’. Models 341350 and 341350-P are kits which include the 34135A and all the necessary accessories.
OPERATING INSTRUCTIONS pH Calibration Calibrate the meter and electrode before each use to ensure accuracy. Connect the pH electrode via the BNC connector to the meter. Open the meter to the fully extended position. Open the battery cover and install a 9V battery.
Page 4
Care of the Conductivity/TDS Cell To prepare the cell, remove any outer protective sheath that may be used in shipping. Cell Storage On sheathed cells, replace the sheath over the cell when storing. For non-sheathed versions, soak the cell tip in de-ionized water for storage. Cell Cleaning After each use, the cell tip should be rinsed with de-ionized water.
pH Troubleshooting Chart Symptom Cause Recommended Solution Long response Clogged Junction Soak in 4.07 M KCL @ 60 C for 30 time or reading minutes drift Strong Alkaline measurement Soak in 0.1 M HCL overnight Deteriorated Gel layer Protein coating on electrode Soak in 1 gm Pepsin dissolved in surface 100ml of 0.1 m HCL for 30 minutes or...