5. Clean the sample
cells with a soft, lint-free
cloth to remove water
spots and fingerprints.
9. Put the sample cell
in the sample cell
holder with the triangle
on the sample cell
aligned with the
reference mark on the
sample cell holder.
Push the lid closed until
a click is heard.
Data management
Show the recorded data
All the recorded data is kept in the data log. There are three types of data logs:
• Reading log—Shows the recorded measurements.
• Calibration log—Shows the calibration history.
• Verification log—Shows the verification history.
1. Push Data Log and select the applicable data log.
2. To show the details of a log entry, select the log entry and then push View Details.
Note: To add a comment to the log entry, push the comments icon.
3. To show only some of the data, push Filter, then select On. The Filter Settings window opens.
4. Select an option.
Option
Description
Time Interval
Selects only the data that was stored during a specific time interval.
Operator ID
Selects only the data that was stored with a specific operator ID.
Sample ID
Selects only the data from the Reading Log that was stored with a specific sample ID.
18 English
6. Apply a small bead
of silicone oil from the
top to the bottom of the
sample cells.
10. Push Read (or
Done if in continuous
mode). Wait for the
instrument to read the
sample.
Note: If auto store is
off, push Options >
Store to save the data.
7. Use the oiling cloth
provided to apply the oil
equally to the surface of
the sample cells.
Remove the excess oil.
Make sure that the
sample cells are almost
dry.
8. Gently and slowly
invert the sample cell to
fully mix the sample. Be
careful not to add air
bubbles.