Specifications; Table 2. Specifications - Fluke 5618B User Manual

Small diameter industrial rtd
Table of Contents

Advertisement

Specifications

See Table 2 for a list of the specifications.
Temperature Range
Nominal Resistance at 0.01 °C
Temperature Coefficient
Drift
[1]
Hysteresis
Sheath Dimensions (Length x Diameter)
Sheath Material
Transition Junction Temperature Range
Response Time
Lead-wire Cable Type
Thermal EMF
Lead-wire Length
Calibration
[1] After 100 hours at maximum temperature.
[2] Temperatures outside this range will cause irreparable damage.
[3] Per ASTM E 644

Table 2. Specifications

Parameter
[2]
[3]
Small Diameter Industrial RTD
Range
–200 °C to 500 °C
(–200 °C to 300 °C for 5618B-6-X)
100 Ω
0.003923 Ω/Ω/ °C
± 0.1 °C at 0.010 °C
± 0.01 °C at 0 °C When Using –196 °C and 420 °C
as the End Points.
5618B-6: 152 mm x 3.2 mm (6 in x 0.125 in)
5618B-9: 229 mm x 3.2 mm (9 in x 0.125 in)
5618B-12: 305 mm x 3.2 mm (12 x 0.125 in)
316 SST
–50 °C to 200 °C
9 Seconds Typical
22 AWG, Teflon™ Insulation
Less than 25 mV at 420 °C
182.9 cm (72 in)
Includes Manufacturer's NVLAP-Accredited
Calibration