Model DMM7510 7½ Digit Multimeter Application Manual
Figure 18: Inductor current linearity test connections
Run the test using the front panel
Running the inductor current linearity test from the front panel:
1. Press the POWER button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select Digi I to select the digitize current function.
3. Swipe to the SETTINGS swipe screen.
4. Set the Sample Rate to 500,000.
5. Set the Count to 50.
6. Press the MENU key.
7. Select Graph to view the graph.
7-12
DMM7510-904-01 Rev. D March 2021