Add 4-port capability upgrade kit

Advertisement

Installation Procedure for the Upgrade
11.Click Yes in answer to the displayed question in the Restart Analyzer? box.
Step 33.Enable Options P04, 419, and 423
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
Option Enable Procedure
1. To start the option enable utility, press UTILITY
An option enable dialog box will appear.
2. Click the arrow in the Select Desired Option box. A list of available options will appear.
3. In the Select Desired Option list, click P04 - 4-Ports.
4. Using the keyboard, enter the license key in the box provided. The license key is printed on the
license message you received from Keysight. Enter this key exactly as it is printed on the
message.
5. Click Enable.
6. Click Yes in answer to the displayed question in the Restart Analyzer? box.
7. Repeat steps 3-6 to enable Option 419, clicking 419 - Src/Rcvr Atten & Bias Ts 4-Port in step
3.
8. Repeat steps 3-6 to enable Option 423, clicking 423 - Combiner & Switches in step 3.
9. When the installation is complete, click Exit.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer's Help menu, click About Network Analyzer.
2. Verify that "P04," "419," and "423" are listed, but "224" and "219" are not listed after
"Options:" in the display. Click OK.
39
If the options have not been enabled or if the option 224 and option 219
licenses have not been removed, perform the prior steps again. If the
options are still not correct, contact Keysight Technologies. Refer to
"Getting Assistance from Keysight" on page
, then
Service
System
3.
Installation Note N5242-90017
, then
Option Enable
.