Fluke 05-450 User Manual

Digital doorway monitor

Advertisement

Quick Links

05-450
Digital Doorway Monitor
Users Manual
PN 4484936
Rev. 1, April 2014
© 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 05-450

  • Page 1 05-450 Digital Doorway Monitor Users Manual PN 4484936 Rev. 1, April 2014 © 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical.
  • Page 3 Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
  • Page 4 To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once every 12 months. Calibration must be done by qualified personnel. Contact your local Fluke Biomedical representative for calibration.
  • Page 5: Table Of Contents

    Table of Contents Title Page Introduction .................... 1 Safety Information ................. 3 Symbols ....................4 General Specifications ............... 5 Driver Option Specification ..............6 Getting Started ..................7 Power Up ................... 7 Radiation Units .................. 8 Checking Parameters ................ 8 Setting Alarm-Points ................
  • Page 6 05-450 Users Manual Cables ................... 17 Optional Remote Alarms ............... 18 Maintenance ..................18 Cleaning Instructions and Precautions ..........18 Replacement of Mains Fuse .............. 18 Detector Connector ................18 Recycling ................... 19 Drawings ....................20...
  • Page 7 List of Figures Figure Title Page Front Panel ....................2 Left Side-Panel ..................7 Alarm Point Notice ..................8 Typical Installation 385x442 ..............20 Detector Enclosure 385x106 ..............21 Detector Mounting 385x120 ..............22...
  • Page 8 05-450 Users Manual...
  • Page 9: Introduction

    Introduction Model 05-450 and 05-450-2200 Digital Doorway Monitor (the Product) is designed to monitor for nuclear radiation. Featuring a wall-mount chassis, the Product has a four-digit LED display that is readable from 9 m (30 ft) away. Backlit indicators warn of low radiation alarm (yellow), high radiation alarm (red), instrument failure (red) and low battery (yellow).
  • Page 10 05-450 Users Manual huk01.eps Figure 1. Front Panel...
  • Page 11: Safety Information

    • The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Fluke Biomedical.
  • Page 12: Symbols

    Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
  • Page 13: General Specifications

    Digital Doorway Monitor General Specifications General Specifications Detectors ..............Two 7.6 cm x 2.5 cm (3 in x 1 in) thick, shielded NaI (T1) scintillation detectors with up to 200 ft cables (NEMA 4x enclosures included.) Display ..............4-digit LED display with 2 cm (0.8 in.) character height Display Range ............
  • Page 14: Driver Option Specification

    05-450 Users Manual Temperature range ..........-20 °C to 50 °C (-4 °F to 122 °F). May be certified for operation from -40 °C to 65 °C (-40 °F to 150 °F) Maximum relative humidity ........Less than 95% (non-condensing) Size Electronics (H x W x D) ..........
  • Page 15: Getting Started

    Digital Doorway Monitor Getting Started Getting Started The Product is designed for ease of use. This section of the manual is designed to help the first-time user get started. Initial power-up and basic features of the instrument are contained in this section. Other sections of the manual provide more detailed information Power Up Plug the power cord into a suitable wall (Mains) outlet.
  • Page 16: Radiation Units

    05-450 Users Manual Please set the alarm point(s) on this insturment to conform to your requirements. The factory-set alarm points may be incorrect for your use. Refer to the instrument manual for more information on setting alarm points. FAILURE TO RESET THE ALARM POINT(S) MAY RESULT IN EXCESSIVE ALARMS OR LACK OF SENSITIVITY.
  • Page 17: Setting Alarm-Points

    Digital Doorway Monitor Getting Started Setting Alarm-Points The LOW ALARM and HIGH ALARM points can only be changed while the instrument is in calibration mode. Switch the top dipswitch CAL MODE (behind the calibration cover) to the right to place the instrument into calibration mode. Changing alarm points is done by holding down the corresponding parameter key and pressing the up or down arrow buttons.
  • Page 18: Operator Controls And Setup

    05-450 Users Manual Operator Controls and Setup Calibration Controls Remove the calibration cover to expose the calibration controls. The calibration controls include the up/down buttons, five calibration potentiometers, and the option dipswitch (detailed in the following subsection). The five potentiometers are detailed below: Warning...
  • Page 19: Dipswitch (Under Calibration Cover)

    Digital Doorway Monitor Operator Controls and Setup Dipswitch (under calibration cover) When the calibration cover is removed, a four-pole dipswitch is accessible that can activate or deactivate options. These four options are: CAL MODE, LATCH ALARM, RANGE, and SINGLE BEEP. •...
  • Page 20: Rs-232 Output

    2 GND pin 5 pin 7 ‘Cable connector has male pins on Model 05-450 side ‘Cable connector has female pins on PC side ‘open up communications with serial port #1 ‘at 2400 bps (baud), no parity, 8 data bits, 1 stop bit ‘no handshaking, buffer size of 8k...
  • Page 21: 9-Pin Data Connector

    Digital Doorway Monitor Operator Controls and Setup 9-Pin Data Connector The 9-pin connector provides output signals from the instrument and input voltage to the instrument. The pin assignments are: pin1- +BATTERY pin2- GND IN pin3- FAIL_L pin4- RS232 DUMP pin5- ANALOG OUT pin6- pin7- HIGH ALARM_L pin8-...
  • Page 22: Common Options And Modifications

    05-450 Users Manual Common Options and Modifications Time and Date Stamp Option Description When an alarm or failure occurs, the Product will print the current reading, date, time, and either ALARM or FAIL to the RS-232 port. The instrument will print once every 30 seconds as long as the alarm or fail condition is present.
  • Page 23: Calibration

    Digital Doorway Monitor Calibration Calibration High Voltage The high voltage is adjustable from 600 V dc to 1200 V dc using the HV potentiometer located under the calibration cover. Ensure that the high voltage is checked only with a high impedance (≥1000 megohm) voltmeter only. A high- voltage checkpoint is located next to the HV potentiometer.
  • Page 24: Analog Output

    05-450 Users Manual Analog Output The analog output is a five-decade logarithmic voltage-out. The maximum voltage-out while under primary power is 6 volts. The maximum voltage out while under battery backup power is 4.5 volts. The five decades are: •...
  • Page 25: Receiving And Installation

    Digital Doorway Monitor Receiving and Installation Receiving and Installation Unpacking Remove the calibration certificate and place it in a secure location. Remove the instrument and accessories (cables, detectors, and other parts) and ensure that all of the items listed on the packing list are in the carton. Check individual item serial numbers and ensure calibration certificates match.
  • Page 26: Optional Remote Alarms

    05-450 Users Manual Note Since the coaxial cable supplies high voltage for detector operation, splicing or re-terminating cables must be done very carefully. Improper termination will result in shorting out the high voltage, a DET FAIL condition and/or blown fuse condition.
  • Page 27: Recycling

    Digital Doorway Monitor Maintenance Recycling The manufacturer of this instrument supports the recycling of the electronics products it produces for the purpose of protecting the environment and to comply with all regional, national, and international agencies that promote economically and environmentally sustainable recycling systems. To this end, the manufacturer strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products.
  • Page 28: Drawings

    05-450 Users Manual Drawings 375-30RWM Typical Installation 375-30rwmtypicalinstallation.eps Figure 4. Typical Installation 385x442...
  • Page 29 Digital Doorway Monitor Drawings M3503_3530_375-30 Enclosure Figure 5. Detector Enclosure 385x106...
  • Page 30 05-450 Users Manual 375-30_3530 Detector Mounting Figure 6. Detector Mounting 385x120...
  • Page 32 05-450 Users Manual...