5. Cleaning and daily disinfecting
7.
If the compressor is dirty, wipe it clean with a
soft cloth moistened with water or mild
detergent.
Caution:
The compressor is not waterproof.
8.
If the outside of the air tube is dirty, wipe it clean with a soft cloth moistened
with water or mild detergent.
9.
Store the parts in the storage compartment in the compressor.
Note: Make sure that the parts are clean and dry. Do not place parts in the
storage compartment if they are still wet.
Store the unit in a clean and dry location.
Note: Nebuliser kit should be replaced after using for 6 months.
5.2

Disinfecting the nebulising parts

Always clean before disinfection. (See Section 5.1)
Always disinfect the medication tank, mouthpiece and other nebulising parts after
the last treatment of the day. To disinfect the parts, use one of the methods
described below:
A. Use a commercially available disinfectant. Follow the instructions provided by
the disinfectant manufacturer.
1.
Submerge the parts in the cleansing solution for the specified period.
2.
Remove the parts and discard the solution.
3.
Rinse the parts with clean hot tap water, shake off excess water and allow
to air dry in a clean environment.
B. Parts may be boiled between 15 to 20 minutes (except for PVC mask).
After boiling, carefully remove the parts, shake off excess water and allow to air dry
in a clean environment.