Calibration (after warming up)
Action
Description
Press [ ON/
1. Power off the scale
OFF ]
Press and
2. While holding the UNIT key, tap the ON/OFF key to turn the scale
Hold [UNIT]
on. The display will show CALu=. The scale is now in calibration
then press
mode.
[ON/OFF]
Choose Unit 3. Press the UNIT key to toggle between weighing units (kg, g, lb, oz).
4. Press ZERO to confirm the weighing unit and proceed to
Press [ZERO]
calibration.
Enter the
5. Using the UNIT and TARE keys, enter the mass value of the
Mass
calibration weight you are going to use:
Press the UNIT key to toggle between digits
Press the TARE key to increment the number
note: The ideal calibration weight is between 2/3 capacity to full
I
capacity in order to get the most accurate results.
6. Press ZERO to confirm the mass value. The display will show a
Press [ZERO]
random number (the AD value).
7. Make sure there is nothing on the scale then press UNIT to
Press [UNIT]
confirm.
8. When the display flashes the calibration mass value, place the
required mass on the platform.
9. Once the weight stabilizes, press the UNIT key to confirm the
Press [UNIT]
calibration adjustment. The display will return to the AD value
10. Remove the calibration weight and turn the scale off. Calibration
is complete!
note: Calibration should be done using accurate weights. The scale will
I
accept the weight you enter, even if it is not correct. Take care to enter the
calibration mass value correctly and only use accurate weights.
All manuals and user guides at all-guides.com