Download Print this page

Calibration; Adjustment - AWS Blade-100 User Manual

Blade series 100g x 0.01g/650g x 0.1g/1000g x0.1g scale

Advertisement

the contents.
4. To return to normal weight display (gross weight),
remove any items from the weighing platform and
press the TARE key again to re-zero the scale.
VII.

Calibration

Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
BL-100
BL-650
BL-1KG
VIII.

Adjustment

1. Press and hold the MODE key until the display
shows "CAL".
2. Press the MODE key again. The display will flash
"CAL" followed by the calibration weight value (see
chart).
3. Place the calibration weight on the center of the
weighing platform. Make sure you are using the
correct weight for your specific model.
Acceptable
Tolerance (±)
0.02g
0.2g
0.2g
Calibration
Weight
100g
500g
500g
Recommended
Accuracy Class
OIML M2
OIML M2
OIML M2