remove any items from the weighing platform and
press the ON/T key again to re-zero the scale.
VII.
Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
CM-500
VIII.

Calibration Adjustment

1. With the scale off, press and hold the MODE and
ON/T keys at the same time until the display shows
a random number then release.
2. Press the ON/T key again and the display will flash
"AL_F" followed by "500.0g".
3. Place a 500g calibration weight (not included) onto
the weighing platform,
4. Once the weight is stable, press the ON/T key once
more and the display will show "AL_F" and return to
normal weighing mode. Calibration is complete.
Acceptable
Tolerance (±)
0.2g
Calibration
Weight
500g
Recommended
Accuracy Class
OIML M2