Troubleshooting - Vatech EzRay Chair User Manual

Table of Contents

Advertisement

5.

Troubleshooting

In instances of abnormal operation, alarm/error messages will be displayed on the
Control Panel. If a problem persists, please request assistance from the customer
support information services.
Alarm/Error Messages
Error Code
E.02
E.03
E.04
A.06
A.07
A.08
EzRay Chair (VEX-S300C) User Manual
A.0X: A problem occurred, and the system performs the
correction automatically. This alarm clears after the correction
is completed.
E.0X: An error occurred. Turn the power off, and then turn it
back on. If the error persists, contact your Service
Representative.
Check Parameter
Error related to X-ray exposure is not
possible to exposure X-ray in the state while
"E.02", "E.03", "E.04", "E.05" where power is
X-ray Generator
maintained. After X-ray exposure related
error occurs, when the equipment is turned
off and turned on, X-ray exposure is usually
performed.
The system needs cooling time due to
continuous operation. (Duty Cycle: 1:60)
X-ray Generator
This alarm clears when the system
temperature goes down to normal.
The Remote Exposure Switch has been held
System
down. Release the switch.
The exposure button has been pressed and
released before the X-ray exposure finished
User
with the default exposure time. Press and
hold the exposure button for the duration of
the exposure time.
5. Troubleshooting
Description
37