accordance with
IEC 61000-4-8
REMARK : U
is the alternating current in the network before the application
r
of the test level.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Tests of Resist-
IEC 60601-1-2:2014
ance to Dis-
Test Level
ruption
Conducted
6 V/m in ISM fre-
high frequency
quency bands be-
disturbance
tween 150 kHz and
variables in ac-
80 MHz
cordance with
IEC 61000-4-6
Radiated high
10 V/m
frequency dis-
80 MHz to 2.7 GHz
turbance varia-
bles in accord-
ance with IEC
61000-4-3
Crystal Imager | Remarks for HF-emission and immunity | 117
Level of
Electromagnetic En-
vironment
Agree-
ment
Use portable and mo-
bile radio sets at a
safe distance from the
device (including the
leads) not closer than
the recommended
protective distance,
which is calculated
according to the equa-
tion suitable for the
RF emission frequen-
cy.
Recommended pro-
tective distance:
6 V/m in
d = 1.2
ISM fre-
quency
bands be-
tween 150
kHz and 80
MHz
10 V/m
d = 1.2
80 MHz to
to 800 MHz
2.7 GHz
3831B EN 20210601 1655
the typical val-
ues as they are
in a
commercial and
clinical environ-
ment.
80 MHz