Calibrating The Touchscreen - AGFA Drystar 5301 User Manual

Hide thumbs Also See for Drystar 5301:
Table of Contents

Advertisement

112 | Drystar 5301, Drystar 5302 | Preventive maintenance schedule

Calibrating the touchscreen

To detect the exact position of your finger on the touchscreen, the touchscreen
must be calibrated. The calibration can be repeated regularly to keep the
touchscreen accurate.
1. Press and hold the text display of the touchscreen for 7 seconds.
Figure 6: Text display
The first calibration screen is displayed.
Figure 7: Calibration screen
2. Press the indicated point on the touchscreen as precise as possible.
The second calibration screen is displayed.
2831F EN 20210601 1655

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Drystar 53025366/1105366/100

Table of Contents