Ultimate Healthcare Heritage II Digital Turn Quick Reference Manual page 2

Alternating mattress replacement system with a cell on cell construction
Hide thumbs Also See for Heritage II Digital Turn:

Advertisement

Mattress Setup
Secure the mattress to the moving parts of the bedframe using
the securing straps, and when the system is fully inflated it is
ready for the patient to be positioned onto the mattress.
Lay the mattress on the bed patient surface ensuring that the
air pipe connector is situated on the bottom left hand side of
the bed foot end (as viewed from the foot of the bed).
Security Straps
There are security straps fitted to the base cover of the
mattress which should be fastened loosely to convenient
points on the bedframe patient surface.
Ensure when fixing the mattress to the bed the security straps
are only fitted to the moving parts of the mattress platform.
Image showing 3 of 6 straps. All 6 straps should be fastened
• The turning function can be activated at any time during the alternating mode of the mattress.
• The function is completed by opening the 'turn valves' on either side of the mattress near the foot section.
• This function is only available during alternating mode and should be deactivated during patient care and transfers.
Valve when mattress is flat
CPR
In the event of a patient suffering a
cardiac arrest and CPR needing to
be administered:
To activate the CPR:
1.
CPR valve is located at the head end on the left hand side
of the mattress.
Gently pull and rotate the dial of the CPR valve to 'click'
2.
into the OPEN position. At the same time, disconnect the
air pipe connector from the Pump to speed up the air
release. The mattress deflates in 10-15 seconds.
To reset CPR:
1.
Ensure the dial of the CPR attachment is rotated until it
'clicks' into the CLOSED position
A Full Risk Assessment should be completed before using this System
Valve showing the LEFT turn symbol
- Please refer to User Manual for guidance
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
Cable Management System Cable management fixings are
located on each side of the mattress underneath the flap of the
cover. The mains power cable should be secured through the
cable management fixing as follows:
• Locate each cable management fixing.
• If necessary, open the press studs.
• Run the mains power cable along the side of the mattress
securing each fixing loop around the cable using the press studs.
Setting up the pump
Whilst holding the pump, unfold the
hanging hooks on the rear of the Pump
and hang it from the bed's footboard.
Ensure power cable the air output
pipe are routed in such a manner so
that they cannot be twisted, trapped, crushed or stressed.
Connecting the Mattress to the Pump
Remove the cover of the air pipe connector and connect the air
pipe connector to the Pump. When turning the power on, the
pump goes directly into auto firm to inflate the mattress. When
fully inflated it reverts to the initial setting of 10 minutes. At this
point the comfort control can be set to the desired weight or
activate the auto detection.
Transport Mode
Transport mode will maintain the cells in their present state for
approximately 48 hours.
Simply disconnect air pipe connector
and place the connector cover over
it. The air pressures in the mattress
will now equalise, but maintain a
degree of comfort.
It is important to restore the mattress
as quickly as possible by reconnecting
the supply tubes to the power unit.
Pull the valve and twist