DISINFECTING
i-STAT Alinity Instrument
Disinfection is recommended between each patient. When instrument is
dedicated to a single patient, disinfect at least once a day. The disinfection
process must begin IMMEDIATELY after the cleaning procedure is
complete. Standard precautions should be taken whenever working with
blood or blood products.
1
Remove a wipe from the
container and squeeze out
any excess fluid before you
begin to wipe the surface.
3
Allow all
surfaces to
remain wet
for 3 minutes.
SENSITIVE AREAS
Avoid forcing liquid into these areas:
i-STAT Alinity Instrument
Cartridge Port
A
10-Pin Connector under the camera
B
C
Gold contacts (2) on the outside of
the battery
i-STAT Alinity — Quick Reference Guide
SECTION 5
2
Wipe all surfaces three times.
This includes front, back, sides,
top and bottom.
4
Art: 731848-01 Rev K
Rev. Date: 02-Apr-2021
3
X
3X
W IPE
W IPE
W
I
P
E
3X
3X
W IPE
W I PE
3
X
Wipe with gauze
until dry.