Note: Be careful not to touch the print head resistor line with
your fingers
10. Clean the print head resistor line.
Gently pass over the resistor line a few times with a lint free cloth, slightly
moistened with Isopropyl alcohol or Ethanol. Do this only in one direction,
i.e. from left to right, without lifting the cloth.
Note: Do not apply any pressure on the print head because
this pressure may cause damage on the interconnections
underneath the print head.
11. Close the print head unit, the hold-down bracket and finally the top cover.
After you have cleaned the print head resistor line and you have closed the
top cover, you will return automatically to the Select calibration menu (see
step 3).
Note: If residue dust is present as part of the cleaning
procedure it will disappear after a few prints.
12. Press the Escape key to return to the Key-operator main menu.