Communication Option
The Communication Option enables digital communication between the UT 150
controller and a PC. It is a factory-installed temperature controller and cable
assembly using an RS-485 connection through a DB9 cable.
This option is supplied with the necessary cable and diagnostic software to set up
and check the connections between the unit and the PC. Follow the steps below to
make the cable connections and to check the data transfer. If you have purchased
the "SpecView Plus Communication Software" with the copy protection key, refer
to the Spec View instructions in parallel with this setup outline.

Cable Installation

1. To install the 25-foot external cable, disconnect the electrical power from both
the unit and PC.
2. Connect the cable end with a black housing to the 9-pin port on the rear of the
Thermo Fisher Scientific.
3. Connect the other cable end with the RS-232/485 Converter to the COM 1 Port
(or other COM port of your choice) on the rear of the PC.
4. If you have purchased the Spec View Plus Communication Software with the
copy protection key, install this key on your parallel port. It may be necessary
to locate (the key between a cable and the parallel port.
5. Apply electrical power to the unit and the PC.
UT150 Communications Setup Parameters
Table shows the default values for UT150 Communications Setup Parameters. To
access these parameter:
1. Hold the SET/ENT button for three second to display the Operating
Parameters. Press and release the SET/ENT button to display the "LoC"
parameter. Assure lower value at "ø". "Press the down arrow to show "-1" in the
lower display and press SET/ENT to acknowledge and enter the Setup Parameter
menu.