Single Setpoint Operation

The following sections describe how to operate the controller in single setpoint
(local) mode. Use this mode when you only need to run the furnace with a specific
setpoint and do not require a programmed sequence of steps - ORA Temperature
change over a period of time.
Setting High Temperature Alarm Setpoint:
1. Press and HOLD for three seconds the "SET/ENT" button to display "modE
rES".
2. Press and release the "SET/ENT" button to display "PrG 0".
3. Press the "UP/RESET" Button to show the lower display the display label "1".
4. Press and release the "SET/ENT" button to select this new value and advance
to the "SSP 25" display, the beginning of the program mode.
5. Press and release the "SET/ENT" button until the High Temperature Alarm
Setpoint value is displayed as "A1".
6. Select an alarm setpoint 10 °C above greater than the target setpoint to be
selected.
7. Press and release the "SET/ENT" button to select this new value in the
controller memory.
8. Press and HOLD for three seconds the "SET/ENT" button to exit this menu.
Accessing Local Mode
1. Press and hold for three seconds the "SET/ENT" button to display "modE rES".
2. Press and release the "UP" button twice to select the display "modE LCL".
Thermo Scientific
Use the Up Arrow button to increase the temperature setpoint
display and change parameter values in the Operation and
Setup Parameter menus. Whenever you change the value of
setpoint or parameter, the decimal point flashes to remind you
to register the changed value with SET/ENT.
Use the Down Arrow button to decrease the temperature set-
point display and to change parameter values in the Operation
and Setup Parameter menus. Whenever you change the value
of a setpoint or parameter, the decimal point flashes to remind
you to register the changed value with SET/ENT.