Using The Advance Function - Thermo Scientific CC58114PC Installation And Operation Manual

Table of Contents

Advertisement

Operation - UT150 Controller

Using the Advance Function

Using the Advance Function
Introduction:
This section describes the sequence to advance or skip segment in a routine
program. This is useful if a power out occurred and the program needs to be
advanced, skipped procedure already computed.
While the program is running, press and hold the "SET/ENT" key for 3 second.
"node" will appear in the upper display. Press the "SET/ENT" key again and "AdV"
will appear in the upper display. "OFF" will appear in the lower display.Press the
"arrow up" key so that "on" with flashing decimal appears in the lower display.
Press the "SET/ENT" key to accept. The controller will automatically return to the
normal display and the increment the program segment by one.
Changing a Program
To make changes only to the target temperature and segment length times for
simple program operation, follow these steps:
1. Assure the indicators beside "RUN" and "L" on the controller display are
extinguished.. If either indicator is illuminated, press and hold the "SET/ENT"
button until the display shows "mode", Select "Res" in the lower display with
the "arrow" buttons. Press and release the "SET/ENT" button once.
2. Press the "SET/ENT" button for 3 seconds to display "mode" in the upper
display and "rES" in the lower display. Press and release "SET/ENT"
repeatedly to display "LoC". Make sure the value below "LoC" is "0" (zero). If it
is not "0", use "DOWN" arrow to select "0" and press and release "SET/ENT"
button to register the change to "0".
3. Press and release the "SET/ENT" button once to show "PrG" on the upper
display.
4. Press the "UP" arrow to make the lower value "1".
5. Press and release "SET/ENT" button twice to display "SP1" Using the arrow
buttons to revise the target setpoint.
6. Press and release "SET/ENT" button twice to display "tml". Using the arrow
buttons to revise the segment time length needed to get to the target setpoint
"SP1".
7. Press and release "SET/ENT" button to display other setpoints and segment
time lengths. Use the arrow buttons to change the temperature set points and
time lengths. Press and release the "SET/ENT" button to register any new
values.
8. Press and HOLD the "SET/ENT" button for 3 seconds to exit the program
menu and return to the reset or standby display.
6-14
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cc58114pbc

Table of Contents