Specifications; Measurement Parameters - IMT Analytics FlowAnalyser PF-300 PRO User Manual

Table of Contents

Advertisement

4

Specifications

4.1

Measurement parameters

IMT Analytics AG
4.1.1
Measuring instrument values
Flow
Range
Accuracy
Ultra-low flow
Range
Accuracy
Volume
Range
Accuracy
Pressure (at flow)
Range
Accuracy
High differential pressure
Range
Accuracy
Low differential pressure
Range
Accuracy
High pressure and
Range
vacuum
Accuracy
Ambient pressure
Range
Accuracy
Oxygen
Range
Accuracy
Temperature
Range
Accuracy
Humidity
Range
Accuracy
1
Standard litres per minute (converted to STP conditions of 21.1°C and 1013 mbar)
*
Tolerance related to the measured value
** Ablsoute tolerance, with steady air flow
*** Non-condensing
Specifications
1
-300 – 300 L/min***
± 1.65 %* or ± 0.04 L/min
(for 10 – 40° C)**
-1 – 1 L/min***
± 1.65 %* or ± 0.01 L/min
(for 10 – 40° C)**
0 – 1000 L
± 1.75 %* or ± 0.10 mL
-50 – 160 mbar
± 0.5 %* or ± 0.1 mbar**
-250 – 250 mbar
± 0.5 %* or ± 0.1 mbar**
-10 – 10 mbar
± 1 %* or ± 0.01 mbar**
-1 – 10 bar
± 1 %* or ± 7 mbar**
500 – 1240 mbar
± 1 %* or ± 5 mbar**
0 – 100 vol %
± 1 vol %**
0 – 50 °C
± 1.75 %* or ± 0.5 °C**
0 – 100 % RH***
± 3 % RH** from 10 % RH to 80 % RH
± 5 % RH** for <10˚% RH and > 80 % RH