8.9
Technical specifications
IMT Analytics AG
Physical data
Dimensions (L × W × H)
Weight
Cable length
Ambient conditions
Operating temperature
Storage temperature
Air humidity (operation)
Air humidity (storage)
Atm. pressure (operation)
Gas
Accuracy
specifications
CO
2
(under standard
conditions)
N
O
2
HAL, ISO, ENF
SEV
DES
Rise times
CO
≤ 90 ms
2
(@ 10 L/min)
N
O, HAL, ISO, ENF, SEV, DES ≤ 300 ms
2
Monitoring
Numerical measurement data and real time graph representation
with FlowLab software.
Deviations from gas setting stated. For example, 50 vol% helium typically reduces
the CO
values by 6 %. This means that a measured mixture consisting of 5.0 vol%
2
CO
and 50 vol% helium is equivalent to a measured concentration of (1
2
vol% = 4.7 vol% CO
.
2
MultiGasAnalyser OR-703
38 × 37 × 34 mm
1.49 × 1.45 × 1.34 inches
<25 g (not including cable)
2.50 m ± 0.1 m
10 – 40 °C, 50 – 104 °F
-40 – 70 °C, -40 – 158 °F
10 – 95 % RH, non-condensing
10 – 95 % RH, non-condensing
525 – 1200 hPa (< 5211 m)
Area
Tolerance
0 – 15 vol%
± (0.2 vol% + 2% of reading)
15 – 25 vol% unspecified
0 – 100 vol%
± (2% vol% + 2% of reading)
0 – 8 vol%
± (0.15 vol% + 5% of reading)
8 – 25 vol%
unspecified
0 – 10 vol%
± (0.15 vol% + 5% of reading)
10 – 25 vol%
unspecified
0 – 22 vol%
± (0.15 vol% + 5% of reading)
22 – 25 vol%
unspecified
8
0.06) * 5.0
–
39