Specifications
4
10
4.1.2
Ventilation parameters
Vti, Vte
Tidal volume of inspiration
and expiration
Vi, Ve
Minute volume of
inspiration and expiration
Ti
Inspiratory time
Te
Expiratory time
Ti/T
Ratio of inspiratory time to
cyc
time of a respiratory cycle
P
Peak inspiratory pressure
peak
P
Mean airway pressure
mean
I:E
Inspiratory/expiratory time
ratio
PEEP
Positive end-expiratory
pressure
Rate
Ventilation rate
PF
Peak flow during
Insp
Inspiration
PF
Peak flow during
Exp
Expiration
C
Static lung compliance
stat
P
Inspiratory pause pressure
plateau
%TP
Post-Inspiratory Pause
"%TP" (=Inspiratory hold
time/Inspiratory time)
Ti
Insp. hold time
Hold
Te
Exp. hold time
Hold
*
Tolerance related to the measured value
** Ablsoute tolerance, with steady air flow
Range
0 – 60 L
Accuracy
± 1.75 %* or ± 0.10 mL
Range
0 – 300 L/min
Accuracy
± 1.75 %* or ± 5 ml/min
Range
0.01 – 60 s
Accuracy
± 0.01 s
Range
0.01 – 90 s
Accuracy
± 0.01 s
Range
0 – 100 %
Accuracy
± 5 %*
Range
0 – 160 mbar
Accuracy
± 0.75 %* or ± 0.10 mbar**
Range
0 – 160 mbar
Accuracy
± 0.75 %* or ± 0.10 mbar**
Range
1:300 – 300:1
Accuracy
± 2 %*
Range
0 – 160 mbar
Accuracy
± 0.75 %* or ± 0.10 mbar**
Range
0 – 2000 bpm
Accuracy
± 1.0 %** or ± 1 bpm
Range
-300 – 300 L/min
Accuracy
± 1.65 %* or ± 0.04 L/min**
Range
-300 – 300 L/min
Accuracy
± 1.65 %* or ± 0.04 L/min**
Range
0 – 1000 mL/mbar
Accuracy
± 3 %* or ± 0.01 mL/mbar**
Range
0 – 160 mbar
Accuracy
± 0.75 %* or ± 0.10 mbar**
Range
0 – 100 %TP
Accuracy
± 0.1 %TP
Range
0 – 60 s
Accuracy
± 0.01 s
Range
0 – 90 s
Accuracy
± 0.01 s