Shodex IC SI-37 4D Operation Manual

Advertisement

Quick Links

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.

1. Introduction

Thank you for purchasing the Shodex product. Shodex IC SI-37 4D is a suppressor type ion chromatography
column for anion analysis, using hydroxide eluents such as potassium hydroxide and sodium hydroxide. It is
suitable for separation of standard anions including fluoride, chloride, nitrite, bromide, nitrate, sulfate, and
phosphate, as well as halogen oxides including chlorite, bromate, and chlorate.

2. Column Components

Please refer to the Shodex website: https://www.shodex.com/en/da/07.html

3. Column Specifications

Product
Code
F6999371
IC SI-37 4D
F6709620
IC SI-90G
Base Material
: Spherical porous particles of polyvinyl alcohol modified with quaternary ammonium
Column Housing : PEEK
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : 10 mM Sodium Sulfate Solution (SI-37 4D)
1.8 mM Sodium Carbonate Solution + 1.7 mM Sodium Bicarbonate Solution (SI-90G)

4. Usable Conditions

Product Name
Recommended
IC SI-37 4D
IC SI-90G
Usable solvents are listed below.
(1) In general, either potassium hydroxide or sodium hydroxide is used.
(2) The standard eluent is 23-mM potassium hydroxide. The higher the concentration of potassium hydroxide,
the faster the elution of ions will be.
(3) Up to 100 % acetonitrile is usable.
Operation Manual
Shodex IC SI-37 4D
Column Size (mm)
Product Name
Flow Rate (mL/min)
Maximum
0.5
0.9
-
-
Resonac Corporation (https://www.shodex.com/)
Particle Size
I.D.
Length
4.0
150
4.6
10
Maximum Pressure
(MPa/column)
20
-
Theoretical Plate Number
(μm)
(Per Column)
3.5
≥ 14,000
9
(Guard Column)
Temperature (°C)
pH
Range
Recommended
3 - 13
30
ver.20221202E
Range
20 - 60
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IC SI-37 4D and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex IC SI-37 4D

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex IC SI-37 4D is a suppressor type ion chromatography column for anion analysis, using hydroxide eluents such as potassium hydroxide and sodium hydroxide. It is suitable for separation of standard anions including fluoride, chloride, nitrite, bromide, nitrate, sulfate, and phosphate, as well as halogen oxides including chlorite, bromate, and chlorate.
  • Page 2 It may lead to abnormal peak shapes, poor separations, and/or low reproducibility. In such cases, please adjust the sample concentration and/or the injection volume. Note · Use of a guard column is recommended to protect the analytical column. Resonac Corporation (https://www.shodex.com/)
  • Page 3 · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. 7.3 Solvent Exchange To replace the solvent, start the system at less than 0.25 mL/min flow rate. Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume. Resonac Corporation (https://www.shodex.com/)
  • Page 4 · When reusing the column stored in a refrigerator, first wait for the column temperature to be ambient, and then remove the end caps and install the column on the system. Resonac Corporation (https://www.shodex.com/)
  • Page 5 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...