DOscan10 Pocket Dissolved Oxygen Tester USER MANUAL Bante Instruments Inc.
Page 2
Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This user manual provides a step-by-step guide to help you operate.
Page 3
If enabled, all of the calibration values and current settings will be deleted or reset to the factory The DOscan10 tester contains 8 menu items in the setup menu, the defaults, the tester must be recalibrated. following table describes the functions of each option.
Page 4
Dissolved Oxygen Calibration Altitude (m) mmHg 101.3 The DOscan10 tester allows 1 or 2 points dissolved oxygen calibration. If you have selected the single point calibration in the setup menu, 100.1 we recommend that you perform a 100% saturation calibration in the.
Page 5
Temperature Calibration Press and hold the Cal key to enter the setup menu. Measurement The DOscan10 tester can be used to measure the water, wastewater, Press the Enter key until the display shows / or / brine and other liquids. If your sample is seawater or water containing large amounts of salt, make sure to set the salinity coefficient before measurement.
Page 6
DOscan10 Preparation of Zero Oxygen Solution Set the barometric pressure and salinity coefficient in the setup menu. Dissolve 500 mg of the sodium sulfate (Na ) reagent and a small Rinse the electrode with distilled water. amount of cobalt (II) chloride hexahydrate (CoCl O) in the 250 ml •...