If you do not press any key within 8 minutes, the tester will switch off automatically to conserve energy. turn clockwise until tight. TDS Calibration The TDSscan10 tester allows 1 point calibration. For better accuracy, we recommend to calibrate the tester regularly. The following table shows acceptable standard solutions for each tester. Keypad...
TDSscan10 Measurement TDS Measurement Rinse the electrode with distilled water, place the electrode into the sample solution and stir gently, make sure that no air bubbles on the sensor surface. Wait for the measurement to stabilize and record the 2. Press and hold the Cal key to enter the calibration mode, the reading.
Fill the distilled water to the mark, mix the solution until the reagent Calculating the TDS Conversion Factor is completely dissolved. The TDSscan10 tester allows setting the conversion factor according to the type of sample. The factory default is 0.50. The following formula Optional Accessories describes the calculation method.
Page 5
TDSscan10 Tester Specifications Model TDSscan10 TDSscan10 L: 0.5 to 100.0 ppm Range TDSscan10 M: 5 to 1000 ppm TDSscan10 H: 0.05 to 10.00 ppt TDSscan10 L: 0.1 ppm Resolution TDSscan10 M: 1 ppm TDSscan10 H: 0.01 ppt Accuracy ± 1% F.S...