Download Print this page

Keysight Y1301A Installation Manual

Diplexer mounting accessory for the m1749b performance mmwave transceiver

Advertisement

Quick Links

Keysight Y1301A Accessory Kit
Diplexer Mounting Accessory for the
M1749B Performance mmWave Transceiver
Installation
Guide

Advertisement

loading

Summary of Contents for Keysight Y1301A

  • Page 1 Keysight Y1301A Accessory Kit Diplexer Mounting Accessory for the M1749B Performance mmWave Transceiver Installation Guide...
  • Page 2 PERFORMANCE OF THIS to the public to use, modify, DOCUMENT OR ANY INFORMATION reproduce, release, perform, CONTAINED HEREIN. SHOULD display, or disclose commercial KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT computer software or WITH WARRANTY TERMS commercial computer software...
  • Page 3 URLs, according to the name of your product: http://www.keysight.com/find/uxm To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL: http://www.keysight.com/find/MyKeysight Information on preventing instrument damage can be found at: www.keysight.com/find/PreventingInstrumentRepair...
  • Page 5 Performance mmWave Transceiver (a remote radio head), a diplexer module is needed, and a mounting bracket is needed to attach the diplexer to the M1749B. The Y1301A kit provides the diplexers, mounting plates, and cables needed to connect the E7515B to two M1749B remote radio heads. This document provides installation instructions for the kit hardware.
  • Page 6: Table Of Contents

    Inspect the shipping container and the cushioning material for signs of stress. Retain undamaged shipping materials for future use, as you may wish to ship the test set to another location or to Keysight Technologies for service. Verify the contents of the container against the table below.
  • Page 7 — Contact the nearest Keysight Technologies office. See “Locations for Keysight Technologies” on page — Keep the shipping materials for the carrier’s inspection. — If you must return a product to Keysight Technologies, use the undamaged original or comparable shipping materials. Installation Guide...
  • Page 8: Overall System Connections

    Overall System Connections Overall System Connections In the diagram of system connections below, the colored paths represent the diplexers and cables which are included in the Y1301A kit. Use this diagram as a reference in setting up the system connections. Figure 0-2 Connections &...
  • Page 9: Installation Procedure

    Picture or Description Notes Torque Wrench Torque-Wrench 1Nm-Torque Part# 8710-2829 8mm-Opening (supplied with Y1301A kit) Torx-screw keys (not supplied) Part# 8710-2726 ESD Precautions Ensure you are working at a static safe workstation before performing the installation process described below. Review the information presented in “Protecting Against Electrostatic Discharge”...
  • Page 10 Introduction Installation Procedure Diplexer Installation Procedure Step Notes 1. Make sure the M1749B is powered off. 2. Remove the serial number plate from the underside of the M1749B by loosening the four screws. (The plate will later be attached to the diplexer bracket.) 3.
  • Page 11 Introduction Installation Procedure Step Notes 5. The bracket, serial# plate, and diplexer should be attached to the underside of the M1749B as shown. 6. Refer to the illustration below in These connectors require careful handling. See making cable connections between “Precautions”...
  • Page 12 Introduction Installation Procedure Step Notes 7. On the two-port side of the Use the short E6610-60005 cables. diplexer, connect the outputs to the LO In and Control inputs of the M1749B. 8. On the one-port side of the Use one of the long E7515-60025 cables. diplexer, connect the input to the LO1 output of the transceiver module on the rear panel of the...
  • Page 13 Introduction Cable Connections Cable Connections Precautions Threaded connectors can be damaged when cables are connected to them, particularly if they are over-tightened (this could break the signal path entirely, making the M1749B non-functional). When attaching cables to threaded connectors, use a torque wrench to apply the proper torque, as described in the table below.
  • Page 14 Introduction Cable Connections The torque wrench should contact the connector fully, with no visible gap. Figure 1-4 Making proper contact between the wrench and the connector When tightening the connection, turn the wrench gently from the far end, to avoid over-tightening. Figure 1-5 Turning wrench from the far end Installation Guide...
  • Page 15 Introduction Cable Connections Be aware that the Power cable has a rectangular locking mechanism on the end of it; once it has been connected to Power on the M1749B, it cannot be removed until you pull back on the locking mechanism as shown below. Figure 1-6 Locking mechanism of Power cable.
  • Page 16 Introduction Cable Connections Connection procedure for cable bundle Use the following procedure to attach cables to the M1749B. The cables in the bundle are labeled for easy identification. Step 1. Turn the M1749B on its side, with the Power connector at the top (this position makes the connectors much easier to reach with a torque wrench).
  • Page 17 Introduction Cable Connections Step 3. Connect the LO input cable to LO In. Step 4. Connect the 9.6 GHz input cable to 9.6 GHz In. Step 5. Replace the 50 ohm terminator onto 9.6 GHz Out. Installation Guide...
  • Page 18 Introduction Cable Connections Step 6. Connect the Rx IF output cable to Rx IF Out. Step 7. Connect the TX IF input cable to Tx IF In. Step 8. Connect the Control cable to Control. Installation Guide...
  • Page 19 Introduction Cable Connections Step 9. Connect the Power cable to Power. Step 10. Return the M1749B back to its normal position (bottom-side down). Installation Guide...
  • Page 20: Protecting Against Electrostatic Discharge

    Introduction Protecting Against Electrostatic Discharge Protecting Against Electrostatic Discharge Electrostatic discharge (ESD) can damage or destroy electronic components (the possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used). Test equipment and ESD — Before connecting any coaxial cable to a test set connector for the first time each day, momentarily short the center and outer conductors of the cable together.
  • Page 21: Service

    Calling Keysight Technologies Keysight Technologies has offices around the world to provide you with complete support for your wireless test set. To obtain servicing information, or to order replacement parts, contact the nearest Keysight Technologies office listed under “Calling Keysight Technologies”...
  • Page 22 Introduction Service Locations for Keysight Technologies For online assistance: http://www.keysight.com/find/assist To contact Keysight Technologies: http://www.keysight.com/find/contactus Alternately, contact the nearest Keysight sales office: Americas Canada Brazil Mexico (877) 894 4414 55 11 3351 7010 001 800 254 2440 United States (800) 829 4444 Asia &...
  • Page 23 This information is subject to change without notice. © Keysight Technologies 2021-2022 Edition 2, March 2022 Y1301-90001 www.keysight.com...