1. On the Counter Navigation screen, push SYSTEM.
2. Select the Custom tab.
3. Select ROTATING BUFFER.

Set the data buffer size

A change to the buffer size causes all current buffer data to be lost and unrecoverable.
1. On the Counter Navigation screen, push SYSTEM.
2. Select the Options and Accessories tab.
3. Select the Data Buffer Size field. Enter a value between 50 and 5000.
4. Push ENTER to confirm.
5. Push YES to clear the data buffer.

Real-time PDF/CSV option

Use this option to send the count data to a PDF file and if necessary to a CSV file. Secure PDF files
are generated in real time or when sample data is selected from the instrument data buffer using the
Export function.

Turn on the PDF option

1. Push SYSTEM.
2. Select the PDF tab.
3. Enter the unlock code and push ENTER.
4. Select the options:
Option
PDF on USB Detect
New Document On
Export Averages
Export to Excel
Warn when USB drive is
X% full

Save the count data to a folder

Count data can be saved to a file with a filename that includes the date and time the file is created, or
a unique identifier.
1. At the particle counter, make a folder to hold the count data. Select Date or enter a name for the
folder.
2. Select Configure File Name.
3. Use up to seven criteria options to make a file name. Options include: area, date, time, location,
user name, serial number or text.
When the file name is configured, the PDF is automatically saved to the folder.
4. Select <None> to remove a previous entry.
Description
Automatically saves the real-time count data to a PDF file when the particle
counter finds a USB memory stick in the USB port.
Sets the criteria to start a new PDF file. New PDFs are made when an area or
location changes or when the date changes.
Includes the average count of a set number of runs and the count data for each
run to a PDF file.
When a PDF file is made, the count data is also transmitted in .csv (comma
separated value) spreadsheet format.
Sets a notification that the USB memory stick is 50–95% full.