No image available after exposure
Details
An exposure is performed and no image becomes available on the
NX workstation.
Cause
1. Due to the automatic exposure control, the dose may become
2. The sensor(s) to trigger the automatic exposure detection of
Solution Check both of following conditions:
1. Check the output of the automatic exposure on the X-ray
2. Check if no objects with a high attenuation are blocking part
too low to trigger the DR detector.
the DR detector are blocked.
console. If the mAs value is low to your experience, make sure
to manually raise the mAs (keep other settings) instead of
repeating in automatic mode.
of the image area.
DR 18M, DR 24M | Problem solving | 63