Dr 18M, Dr 24M - AGFA DR 18M User Manual

Hide thumbs Also See for DR 18M:
Table of Contents

Advertisement

DR 18M, DR 24M

Electrical connection DR Detector
Rated power supply
(power over ethernet)
Power consumption
Environmental conditions (during normal operation)
Room temperature
Maximum temperature change rate
Humidity (non condensing)
Atmospheric pressure
Environmental conditions (during storage)
Temperature (ambient)
Maximum temperature change rate
Humidity (non condensing)
Atmospheric pressure
Warming-up time
5 minutes minimum
10 minutes recommended
30 minutes before calibration
Dimensions
Dimensions
width x length x height
Weight
Chest wall distance
DR 18M, DR 24M | Technical Data | 67
12V DC
30 W
between +5 °C and +35 °C
max. 10 °C/hour
between 20% and 75% RH
(non-condensing)
between 500 hPa and 1060 hPa
between -10 °C and +40 °C
max. 15 °C/hour
between 5% and 95%
(non-condensing)
between 500 hPa and 1060 hPa
DR 18M: 267.5 x 194.5 x 14.2 mm
DR 24M: 327.5 x 253.7 x 14.2 mm
DR 18M: 0.92 kg
DR 24M: 1.20 kg
less than 2.0 mm
0380A EN 20180703 1429