3.5.6 Is Start, Is Interval

Both the parameter ranges for Is Start and Is Intervals are 0.0~1000.0 msec, these
parameters are for the surge current measurement.
Picture shown below indicates the measurement of the surge current start at Is Start after
output enabled, and measurement time is Is Interval:
For example:if Is Start = 500ms, Is Interval = 200ms, then Is measurement shown on main
manual is the maximum current measured between 500~700ms after output enabled。
Following processes demonstrate how to set Is Start = 500ms and Is Interval = 200ms:
Save
1. Press
35
│ www.deltaww.com
Current
Is
Is Start
Recall
Save
Recall
or
to move cursor to Is Start
S E T U P
V
L i m i t
I
L i m i t
O C P
D e l a y
V o l t a g e
V o l t a g e
D e g r e e
D e g r e e
I s
S t a r t
I s
I n t e r v a l
P r e s s
S E T U P
V
L i m i t
I
L i m i t
O C P
D e l a y
V o l t a g e
V o l t a g e
D e g r e e
D e g r e e
I s
S t a r t
Is Interval
:
1 5 0 . 0
:
:
L e v e l :
1 5 0 V
S e n s e :
V s e n s e
O n
:
1 0 0 . 0
O f f
:
2 0 0 . 0
:
:
t o
S w i t c h
:
1 5 0 . 0
:
:
L e v e l :
1 5 0 V
S e n s e :
V s e n s e
:
O n
1 0 0 . 0
:
O f f
2 0 0 . 0
:
5 0 0 . 0 m s