User Manual
A9000
AC Power Source
4. By pressing numerical key or rotating knob, change parameter to 1.300, press
confirm the setting
3.6.2 GPIB, RS232 Communication Setting
DME-A902ABT SERIES provide remote control operation, there are GPIB or RS232
interface for selection. Only one interface can be activated at the same time:
Operation mode
Manual(front panel)
Remote(GPIB)
Remote(RS232)
How to switch manual to remote control:Send any command to AC source directly, AC
source enter remote control mode automatically.
Two ways to Switch from remote to local control: 1. Stop remote control, press
longer than 2 seconds until "SET MODE" shown on upper left on the screen; 2. Send
*CLS command.
Switch remote control between GPIB or RS232:Back to manual operation as
demonstrated above, then use another communication interface to do the control.
49
│ www.deltaww.com
W a v e
A
W a v e
B
A D D R : 3 0
G P I B
B a u d R a t e : 9 6 0 0
R S 2 3 2
R S 2 3 2
P a r i t y
R e m o t e
I n h i b i t : O F F
P o w e r
O n
O u t p u t
P h a s e
S e l . : T h r e e ( S Y N . )
C O N F I G
W a v e
A
W a v e
B
G P I B
A D D R : 3 0
B a u d R a t e : 9 6 0 0
R S 2 3 2
R S 2 3 2
P a r i t y
R e m o t e
I n h i b i t : O F F
P o w e r
O n
O u t p u t
P h a s e
S e l . : T h r e e ( S Y N . )
Controller
Manual
(front panel)
: D S T 2 0
: C S I N
C F : 1 . 2 0 0
→
: N O N E
S t a t e
: O F F
: D S T 2 0
: C S I N
C F : 1 . 3 0 0
→
: N O N E
S t a t e
: O F F
Remote
(GPIB)
○
╳
╳
Remote
(RS232)
╳
○
╳
to
Enter
╳
╳
○
–
/
Unlock