RBC Bioscience MagCore HF48 Maintenance Manual page 67

Automated nucleic acid extractor
Hide thumbs Also See for MagCore HF48:
Table of Contents

Advertisement

Issued By: RBC Bioscience Corp.
5. Screw off the 4 screws (3mm hex wrench) on the base of the Shutter motor and the 2
screws (2mm hex wrench) on the timing wheel.
6. Install a new Shutter motor.
7. Attach the timing belt on the Shutter motor.
8. Push the Shutter motor backward and fasten the screws on the side of the Shutter
motor.
9. Attach the sheet metal cover on the machine.
10. Plug in the power plug of the Shutter motor.
11. The replacement of the Shutter motor is finished.
MagCore HF48 Service Manual