Installations - RBC Bioscience MagCore HF48 Maintenance Manual

Automated nucleic acid extractor
Hide thumbs Also See for MagCore HF48:
Table of Contents

Advertisement

Issued By: RBC Bioscience Corp.
Installation
Place the system in a suitable location and use the leveling glides to
fine-tune the elevation. Remember to tighten the screws.
Next Step
1
Rotate and pull out the knobs on both sides of the top cover of part
chamber using coins.
Next Step
Open the top cover of part chamber.
Please note the weight of the top cover of
part chamber. Two people are required to
lift it up.
Next Step
Cut the fixation belt and close the top cover of part chamber.
Next Step
MagCore HF48 Service Manual
2
Revised: February,2014
Cut the fixation belts on the racks.
Next Step
Plug in the power cord and turn on the main power.
Please make sure that the voltage of the
outlet is between 220V and 240V. Keep
your hands dry. Please also make sure
that the machine is turned off when you
insert the plug.
Next Step
Turn on the power and the system will start initializing setup
automatically.
!
Please set the local time (refer to page 24) for
the first installation, and perform the "Piston
Test" (page 20).
Please finish reading the operation manual
before operating.
Ver. 2014-2
RBCBioscience
9