Issued By: RBC Bioscience Corp.
Test home value
After installation, it is necessary to make sure that the position of modules and axes
have not been shifted during shipping. Please check if the values conform to the given
values in the original factory document.
1. Please make sure that the instrument is turned off.
2. Disconnect the brake from the instrument and install the electronic brake thread.
3. Install Z-axis calibration tool on Operating Area A and manually rotate the Y-axis
belt anti-clockwise to lower the Y-axis till it reaches the home calibration tool
a n d c o r r e c t l y i n s e r t t h e
c a l i b r a t i o n p o i n t i n t h e
calibration tool.
18良距離
Rotate the Y-axis belt anti-
clockwise to lower the
Y-axis.
4. Repeatedly move the Y-axis up and down to make sure whether the calibration point
can insert into or separate from the Z-axis calibration tool without any difficulties.
5. Prepare a computer and connect it to the instrument. Turn on the instrument.
MagCore HF48 Service Manual
Put the Z-axis calibration tools at the disposition
holes on the costal plates of the Cartridge Racks.
Revised: February,2014
Ver. 2014-2
6
RBCBioscience
11