Piston Test - RBC Bioscience MagCore HF48 Maintenance Manual

Automated nucleic acid extractor
Hide thumbs Also See for MagCore HF48:
Table of Contents

Advertisement

Issued By: RBC Bioscience Corp.

Piston test

Maintenance
UV Sterilization
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Please select Item
Select Piston test in the Maintenance menu.
Next Step
Piston Test
Press START to run
Please Place the"system test tools"in advance in Operating areas
and put 1ml water in all Sample Tubes. Press Start to begin the test.
Next Step
O-Ring & Piercing Needles
Piston Test Step 1
Replace O-Rings
Machine taking up Tips, please check whether the Tips move smoothly. If you observe a Tip Holder
stuck, please press the STOP button immediately to stop and remove the jammed Holder.
Next
Select item and Press Start to run
The purpose of this step is mainly to test whether the instrument
can withdraw tips smoothly.
The display shows Piston Test Step 1, please note that the
illustration shown on screen is only an example. Please check if the
instrument can withdraw tips smoothly. If yes, press Next. If you
find any tips stuck in its holder, press Abort to stop the test and
replace the tip/tip holder. Run the test again and if the problem
persists, contact your local distributor for technical support.
Next Step
MagCore HF48 Service Manual
Moving Test
Heater Test
Rejection Tip
Update Program Files
Back
Start
Stop
B
B
A
A
E
D
Back
Clean Piercing Needles
Start
Stop
Abort
Back
Revised: February,2014
O-Ring & Piercing Needles
Piston Test Step 2
Replace O-Rings
Tip will absorb water from Sample Tube, please check the liquid level of each columns are
at same horizontal.
Select item and Press Start to run
The purpose of this step is to test if there is any leakage from the
tips.
The system will absorb water from Sample Tubes. Please check
if the liquid level is the same across all tips. If yes, press Finish. If
the liquid level is not the same throughout, please press Finish
first, then replace the O-ring of the leaking tip and test again. If
the problem persists, please contact your local distributor for
technical support.
Ver. 2014-2
Clean Piercing Needles
Start
Stop
Finish
Back
RBCBioscience
20