Connection Of Plugs To The Main Cpu Board, D1 - Siemens MOBILETT Plus Installation Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 14

Connection of plugs to the main CPU board, D1

1.
Remove the protection plate of the CPU board below to expose the main CPU D1
(see Fig. 18).
2.
Disconnect pin plug K9 from board D1 (see Fig. 18).
3.
Place the disconnected K9 cable in between the chassis frame and the D4 board
(see Fig. 15).
4.
Mount a flat tie holder according to pos. 1 / Fig. 15.
5.
Place a cable tie (2.5x100mm) around the disconnected K9 cable according to
pos. 2 / Fig. 15.
6.
Place the large shrinking tube (Part No. 66 11 115, 50 mm length) around the
loose K9 pin plug (see pos. 3 / Fig. 15).
7.
Connect pin plug K9 to K9b of cable 65 61 992 (see Fig. 15).
8.
Shrink the shrinking tubing (see pos. 3 / Fig. 15) over the connected pin plugs K9
and K9b.
MOBILETT Plus/ -E/ -HP
remote control
system
1
Fig. 15 Positioning of cable tie and shrinking tubing
SPR8-220.031.04
Rev. 04
D4
Page 14 of 30
09.03
CS PS 24
Installation
2
3
2
Siemens AG
Medical Solutions