Installation
16.
Cut away the shrinking tubing at pos. 1 / Fig. 12.
CAUTION
There are three twisted pair of cables inside: a white and green cable, a white and
blue cable, and a white and purple (violet) cable.
17.
Separate the twisted white and blue cable, and the twisted white and purple cable.
18.
Shrink the twisted white and green cable using a shrinking tube (65 82 407) with
the others that were there before (leave out the twisted white and blue, the yellow
and white, and the twisted white and purple cables).
19.
Reconnect the cables as before, see pos. 3 (orange), 4 (2 x white) and 5 (brown) /
Fig. 11.
20.
Cut the shrinking tubing of the twisted white and blue cable and the twisted white
and purple cable and strip the cables.
NOTICE
21.
Strip the six cables at pos. 2 / Fig. 12 approximately 10 mm.
Siemens AG
Medical Solutions
Do not cut the cables!
If you accidentally cut the cables, they must be spliced.
Be sure to cut the shrinking tubing only, not the cables.
1
2
Fig. 12 Cutting of shrinking tubing and striping of cables
Do not shorten the free ends! Strip the cables carefully, approxi-
mately 10 mm .
SPR8-220.031.04