4180, 4181
Safety Information
Safety Specifications
Complete specifications are in the Operators Manual.
Temperature range
(@ 23 °C ambient, 0.95 emissivity)
Power
Size (HxWxD)
Weight
Ambient temperature
Ambient relative humidity
Altitude
Fuse(s)
Safety
Electromagnetic Compatibility (EMC)
4
4180
-15 °C to 120 °C
115 V ac (±10%),
50/60 Hz, 250 W
230 V ac (±10%),
50/60 Hz, 250 W
9.1 Kg (20 Ib)
5 °C to 35 °C (41 °F to 95 °F)
maximum 80% for temperature <31 °C, decreasing linearly to 50 % at 40 °C
115 V: T 6.3 A, 250 V
230 V: T 3.15 A, 250 V
IEC 61010-1: Overvoltage Category II, Pollution Degree 2
International................................. IEC 61326-1: Basic Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-
coupled radio frequency energy that is necessary for the internal function of the
equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic
and those directly connected to a low-voltage power supply network that
supplies buildings used for domestic purposes. There may be potential
difficulties in ensuring electromagnetic compatibility in other environments due to
conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and
may not provide adequate protection to radio reception in such environments.
Korea (KCC) ............................... Class A Equipment (Industrial, Broadcasting, &
Class A: Equipment meets requirements for industrial electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and not to be used in homes.
USA (FCC) .................................. 47 CFR 15 subpart B. This product is considered
35 °C to 500 °C
115 V ac (±10%),
50/60 Hz, 1100 W
230 V ac (±10%),
50/60 Hz, 1100 W
356 x 241 x 216 mm
(14 x 9.5 x 8.5 in)
9.5 Kg (21 Ib)
<2000 meters
115 V: F 10 A, 250 V
230 V: F 5 A, 250 V
IEC 61010-2-010
Communication Equipment)
an exempt device per clause 15.10
4181