Troubleshooting
Error – Cause –
Remedy
54
6.2 Error – Cause – Remedy
6.2
Error – Cause – Remedy
6.2.1
6.2.1

Pop-up message

Pop-up message
Error
Error
` Possible cause
Communication
`
One or more
error
VACUU·BUS components
were removed.
Frequency con‑
`
Address incorrectly con‑
verter (FC) error
figured.
`
Temperature too high.
FC defective.
`
Error
`
Valve defective.
affecting control
Error at pump
`
Check VMS‑B (switching
device).
Error at digital I/O
`
No power supply at IN of
module
I/O module.
Plug pulled out.
`
`
An error occurred in the
system and the I/O mod‑
ule relayed it to the con‑
troller.
Error at analog I/O
No power supply.
`
module
Error
Ambient temperature too
`
affecting Peltronic
high, device overheating.
`
Very high condensing
capacity.
Device defective.
`
Error at vacuum
Vacuum sensor defective.
`
sensor
Overpressure
`
Pressure too high.
`
Measuring range
exceeded.
Underrange
`
Pressure below measur‑
ing range.
`
Vacuum sensor calibra‑
tion incorrect.
 Remedy
Deactivate relevant
VACUU·BUS compo‑
nents.
Perform component
recognition.
Configure correct
address.
Replace defective com‑
ponents.
Check address.
Replace defective com‑
ponents.
Send in defective
device.
Connect power supply.
Check plug‑in connec‑
tion.
Remedy cause of exter‑
nal error.
Connect power supply.
Eliminate reason for
Peltronic overheating.
Send in defective
device for repair.
Replace defective
device.
Send in defective com‑
ponent.
Acknowledge warning.
Eliminate cause of
overpressure.
Calibrate vacuum sen‑
sor correctly.
20901531_EN_VACUU·SELECT_V1.0_010621
Personnel
Specialist
Resp. spe-
cialist
Specialist
Resp. spe-
cialist
Specialist,
resp. spe-
cialist
Specialist
Specialist
Resp. spe-
cialist
Operator,
specialist
Specialist