C- OPERATIONAL TROUBLESHOOTING
PROBLEM
Fails to respond to touch
selection of a key
Unit fails to turn on when
"ON/OFF" switch is set to on.
Unit
hangs
with
Bio-Med
Devices, Inc. on the display.
Internal
regulator
output
pressure varies
Low internal pressure regulator
output
System pressure reads zero
Incorrect peak pressure and
PEEP/CPAP reading
Uncontrolled
auto-cycle;
remains in inspiratory mode
Peak pressure high limit alarm
activates
Peak Pressure Low limit alarm
activates
Ventilator does not start
Crossvent-2i+ fails to operate
on battery
M E N U
A N D
T R O U B L E S H O O T I N G -
POSSIBLE CAUSE
1. Keys locked
2. Defective touchscreen key
1. Battery low and plug-in power supply
not connected.
2. No power at outlet and battery low.
Initialization error.
Low pressure/flow from gas source; fault in gas
supply
1. Output pressure not set; insufficient gas
supply
2. Dirty or clogged filter
1. Peak pressure limit control is fully off
2. Tube disconnected between patient and
ventilator
1. Pressure transducer disconnected or
malfunction
2. Pressure transducer is out of calibration
1. Flow Trigger setting is too sensitive
2. Pressure transducer malfunction
1. Accumulation of secretion
2. Change in compliance; blockage in
airway or tubing
3. Patient tries to breathe independently
4. Endotracheal tube has slipped down
into a bronchus
5. Change in patient positioning
6. Pneumothorax
7. Sigh
1. Disconnected patient circuit
2. Endotracheal tube has slipped down
into a bronchus
3. Water in patient circuit
4. Decrease in lung compliance
1. Power cable not connected and battery
discharged
2. Low battery and no power in electrical
outlet
3. Blown fuse in plug-in power supply
4. ON/OFF switch in the OFF position
1. Low battery
2. Missing battery
3. Defective battery
C O N T .
CORRECTIVE ACTION
1. Press LOCK key twice to unlock
2. Submit for service
1. Plug in charger to active outlet and
charge battery
2. Plug into active outlet and charge
battery
Submit for service.
Increase gas volume/pressure
1. Set output pressure; check gas
2. Submit for service.
1. Adjust the PIP limit control
2. Connect tube
1. Submit for service.
2. Submit for service.
1. Increase Flow Trigger setting
2. Submit for service.
1. Examine patient
2. Remove the blockage or replace
the airway/tube
3. Re-evaluate patient status and vent
settings
4. Correct the tube position
5. Re-evaluate
settings
6. Treat patient
7. Set the high alarm limit higher
1. Reconnect patient circuit
2. Reposition endotracheal tube
3. Empty water from patient circuit
4. Re-evaluate patient
1. Connect power cable; charge or
replace battery
2. Change outlet; charge or replace
battery
3. Replace transformer
4. Switch to ON
1. Charge battery
2. Insert and attach battery
3. Replace battery
31