ResMed VPAP III ST-A User Manual page 23

With quicknav
Hide thumbs Also See for VPAP III ST-A:
Table of Contents

Advertisement

Available languages

Available languages

Dimensions (L x W x H):
Weight:
Air outlet:
Pressure measurement:
Flow measurement:
Power supply:
Housing construction:
Environmental
conditions:
Electromagnetic
compatibility
Air filter:
Air tubing:
IEC 60601-1
classifications:
This flow generator is not suitable for use in the presence of a flammable anaesthetic mixture
with air, or with oxygen or nitrous oxide entrained in the flow generator airpath.
Displayed values
Value
Pressure sensor at air outlet
Pressure
Flow sensor in flow generator*
Leak
Tidal volume
Respiratory rate
Minute ventilation
*
Results may be inaccurate in the presence of leaks or supplemental oxygen.
** The displayed values are estimates. They are provided for trending purposes only.
Results may be inaccurate if the tidal volume is below 50mL.
Notes:
The manufacturer reserves the right to change these specifications without notice.
Further clinical information is available through your authorised ResMed service centre.
270 mm x 230 mm x 141 mm
2.3 kg
22 mm taper, compatible with ISO 5356-1:2004 Anaesthetic &
Respiratory Equipment - Conical Connectors
Internally mounted pressure transducer
Internally mounted flow transducer
AC 100–240V, 50–60Hz, 2.2A; AC 110V, 400 Hz, 2.2A;
DC 24V, 2A
Flame retardant engineering thermoplastic
Operating Temperature: +5°C to +35°C
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -20°C to +60°C
Storage and Transport Humidity: 10%–95% non-condensing
Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to IEC60601-1-2,
for residential, commercial and light industry environments. For
further details, see "Guidance and Manufacturer's Declaration -
Electromagnetic Emissions and Immunity" on page 19.
Two-layered, powder-bonded, polyester non-woven fibre
Flexible plastic, 2 m or 3 m length
Class II (double insulation)
Type CF
Continuous operation
Range
Accuracy
-5 to 30 cm H
O
±0.5 cm H
2
measured value)
0–120 L/min
50–3,000 mL
6–60 BPM
±0.5 BPM
0.6–60 L/min
O
(+ 4% of
2
**
**
**
Display Resolution
0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min
System Specifications
17