Transport, transporting and parked
positions
Transport of the Leica M844 F40 and Leica M820 F40
Warning 9
Beware of:
• Uncontrolled lateral movement of the swing arm!
• Tilting of the stand!
• Feet in lightweight shoes could become trapped
beneath the casing of the base.
➩Before transport, always set the Leica M820 F40 and
Leica M844 F40 surgical microscopes to the
transport position.
➩Never move the stand in the extended condition.
➩Always push the instrument to displace it; never pull it.
➩Never roll over cables lying on the floor.
Caution 3
Surgical microscope can move without warning!
➩Always lock the footbrake when you are not moving
the system.
Transport position
Whenever you transport your Leica M820 F40 and Leica M844
F40, first bring it into transport position.
Caution 2
There is a risk of damage to the Leica M820 F40 and
Leica M844 F40 surgical microscopes from
uncontrolled tilting!
➩Firmly grasp the handles before triggering the
"All Brakes" function.
➩Unplug and secure the power cable.
➩Pull the stop lever and bring it into a vertical position.
➩Grasp and turn one or both handles to release the brakes
(All Brakes).
➩Move the swing arm up and down until the transport lock
engages.
➩Bring swing arm into transport position.
➩Release the handle.
➩Turn the control unit towards the swing arm.
➩Hang the footswitch on the suspension device.
➩Step on the footbrake release lever to release the footbrakes.
➩Move the Leica M820 F40 and Leica M844 F40 by the handle.
Stop lever
Handle
Footbrake and release lever
Leica M844 – Leica M820 / Ref. 10 713 294 / Version I
Ensure that the display of the control unit does not col-
lide with the XY unit!
Transport of the Leica M844 F19 and Leica M820 F19
Warning 9
Beware of:
• Uncontrolled lateral movement of the swing arm!
• Tilting of the stand!
• Feet in lightweight shoes could become trapped
beneath the casing of the base.
➩Before transport, always set the Leica M820 F19 and
Leica M844 F19 surgical microscopes to the
transport position.
➩Never move the stand in the extended condition.
➩Always push the instrument to displace it; never pull it.
➩Never roll over cables lying on the floor.
Caution 3
Surgical microscope can move without warning!
➩Always lock the footbrake when you are not moving
the system.
Transport position
Whenever you transport your Leica M820 F19 and Leica M844
F19, first bring it into transport position.
➩Unplug and secure the power cable.
➩Pull the retaining pin and engage the retaining hook.
➩Release the articulation brakes.
➩Bring swing arm into transport position.
Articulation brakes
Handle
➩Tighten the articulation brakes.
➩Turn the control unit towards the XY unit.
➩Place footswitch in the carrier.
➩Release the foot brakes by depressing the foot lever.
➩Move the Leica M820 F19 and Leica M844 F19 by the handle.
Preparation for operation
Retaining hook
35