5. When starting and finishing the BLG-TestProgram, perform the following procedure.
• Starting the BLG-TestProgram
Procedure of Starting
(1) Mount a lens to be tested to the lens mount block.
(2) Connect the D-sub 15-pin cable to the lens mount block, and
turn on the B-L Control Box.
(3) Start "BLG-TestProgramV[][][].exe".
(4) Click [Open] to connect communication with the B-L Control
Box.
Note: When the B-L Control Box or personal computer to be used
is started for the fi rst time, a password must be entered.
Enter a password and click [OK], then proceed to step 5.
Connection to the lens mount block is established and the
[Power] button is made active.
When the password input window opens, ask service head-
quarters about password. In this case, the ID shown in
"COM/USER Characteristic Number" must be informed
("YdDGAAHTJ" in the example above).
Since this ID number is unique to the B-L Control Box and
personal computer, if used another B-L Control Box or per-
sonal computer, another password is necessary.
(5) Connection to the lens mount block is established and the [Power]
button is made active.
• Finishing the BLG-TestProgram
Procedure of Finishing
(1) Click [Power] to turn off the power to the lens.
(2) Confirm that the two LEDs on the B-L Control Box are not lit.
B-L Control Box
(3) Click [Close] to disconnect communication between personal
computer and B-L Control Box.
(4) Click [Finish] to exit the program.
5-7
SEL24105G
LEDs for power supply check
– Continued on next page –