EDAN Acclarix LX9 Series User Manual page 107

Diagnostic ultrasound system
Table of Contents

Advertisement

Acclarix LX9 Series Diagnostic Ultrasound System User Manual
4. Prepare the disinfectant solution (Cidex OPA 0.55% or Cidex Glutaraldehyde 2.4%). Refer to the
instructions provided by the disinfectant manufacturer for the concentration of the disinfection
solution, method of dilution, method of disinfection, temperature and cautions during use.
5. Place the cleaned and dried transducer in contact with the disinfectant (refers to figure 6-3 for the
contacting area) for the time specified by the disinfectant manufacturer. For example, the contact
time recommended by the manufacturer for soaking in Cidex Glutaraldehyde(2.4%) is at least 45
min.
6.
Rinse the transducer thoroughly with sterile water to remove all chemical residues. For example,
it is required to flush the transducer after soaking in Cidex Glutaraldehyde(2.4%) with plenty of
sterile water (about 2 gallons) at least one time. Or, follow the complete rinsing instructions
provided by the disinfectant manufacturer to rinse the transducer.
7.
Wipe off the water on the transducer with a dry sterile cloth. Leave the transducer to air dry.
8. Inspect the transducer to ensure that there is no damage.
Disinfectant
Figure 6-3 Depth of the Transducer Immerged into Disinfectant
1. Do not immerse the transducer connector. If the cable connector is immersed, do not plug the
connector into the system. Rinse the connector under running water and dry it thoroughly. If
necessary, contact EDAN for service.
2. Prohibit infiltration of any type of liquid into the device or the transducer.
3. Do not immerse the cable and connector of the transducer into solutions. Transducers can be
submerged to, but not including, the strain relief of the transducer array. Do not immerse or soak
any part of a transducer in any cleaning material not listed in the recommended list of disinfectant.
4. Use the immersion method to disinfect the intra-cavitary and intra-operative transducers.
5. Only non-immersion method can be used with solution of ethanol or isopropanol..
L17-7SQ
E10-3HQ
E8-4Q
E10-3BQ
WARNING
- 98 -
Transducers and Biopsy
ECL12-3Q